The Transmille EA002 Clamp Coil Adapter is designed for the calibration of both wound (AC only) & Magnetic field (hall effect) AC/DC clamp meters the current coil offers several unique features built in a robust construction. The coil is designed to immerse both jaws of the clamp meter in the magnetic field while allowing the ‘gap’ or opening where magnetic flux will escape to protrude through the coil and stay out of the strongest part of the field. This makes the reading less dependent on the position of the clamp within the coil and also the quality of the jaw closing, allowing greater confidence in the calibration. Specifications.
Connection to Calibrator: Via supplied 4mm to 4mm connection leads (x2) to 3000 series current terminals. Coil Configuration: 2 Turn (LHS), 10 Turn (RHS), 50 Turn (CENTRE). Coil Type: High accuracy balanced configuration. Min. internal jaw dimensions: 10mm (2 Turn & 10 Turn), 25mm (50 Turn). Maximum Current: 30A. Maximum RMS voltage: 4V. Frequency Range: DC to 500Hz.
2 Turn Coil Accuracy (Input 0 to 30A : Freq. DC - 30Hz to 60Hz : Effective Output 0 to 60A). Effective accuracy - Coil only (wound clamps, 90 Day Rel.) 0.35% + 0.008A. Effective accuracy - Coil only (hall effect clamps, 90 Day Rel.) 0.48% + 0.07A.
10 Turn Coil Accuracy (Input 0 to 30A : Freq. DC - 30Hz to 60Hz : Effective Output 0 to 300A). Effective accuracy - Coil only (wound clamps, 90 Day Rel.) 0.41% + 0.01A. Effective accuracy - Coil only (hall effect clamps, 90 Day Rel.) 0.59% + 0.11A.
50 Turn Coil Accuracy (Input 0 to 30A : Freq. DC - 30Hz to 60Hz : Effective Output 0 to 1500A). Effective accuracy - Coil only (wound clamps, 90 Day Rel.) 0.24% + 0.04A. Effective accuracy - Coil only (hall effect clamps, 90 Day Rel.) 0.45% + 0.42A.
|