The Ando AQ6319 is an optical spectrum analyzer which can measure loss wavelength characteristics and propagation characteristics on optical fiber cables, optical filters and the like, in addition to spectral measurements on light sources such as LDs and LEDs. The measurement wavelength range is wide at 600 to 1700 nm, enabling optical spectrum analysis in the near infrared range. In addition to basic features such as high resolution, high sensitivity, high accuracy, wide dynamic range and superior linearity, the AQ6319 has a wide variety of functions such as program measurement functions and numerous data processing functions, including analysis functions and template functions. In terms of data output capabilities, the AQ6319 can provide screen hard copies through the internal high-speed printer or an external printer. In addition, waveforms and programs can be read from and written to the internal hard disk and floppy disk, or read/written through the LAN interface. The AQ6319 also includes GP-IB and RS-232C interfaces as standard features, enabling fully remote control. Lastly, the AQ6319 has an intuitive graphical user interface using a keyboard and mouse for user-friendly operation.
|
|
The Deviser E8600 is a high-precision diffraction-grating, high-resolution optical spectrometer with wavelength range of 600 nm to 1700 nm. The 10.1” LCD touchscreen and concise graphical user interface of AE8600 offer the easiest way to handle optical spectrum analysis. It provides a wide selection of test methodology, including laser spectrum scans (DPB, FP), WDM system testing, EDFA system testing, transmittance and drift testing, which are essential for in-field and factory applications. The AE8600 offers exceptional stability and reliability, high-speed spectral sweeping, and multiple ways to export and analyze measurement data. It’s the ideal instrument for fast and precise optical spectral testing to satisfy long-term investment with the best cost performance value. Specifications.
Optical Spectrum Measurement. Wavelength range: 600 to 1700nm. Wavelength resolution bandwidth: 0.02 to 2nm. Wavelength resolution setting: 0.02nm, 0.06nm, 0.1nm, 0.2nm, 0.5nm, 1nm, 2nm. Wavelength accuracy; 1520 to 1620 nm ±0.02 nm, 1450 to 1520 nm ±0.04 nm, Entire wavelength range ±0.1 nm. Wavelength repeatability: ±0.005 nm (1 minute). Wavelength linearity; ±0.01 nm (1520 to 1580 nm), ±0.02 nm (1450 to 1520 nm, 1580 to 1620 nm). Minimum sampling resolution: 0.001nm. Applicable fiber: SM(9.5/125μm), MMF(50/125μm, 62.5/125μm).
Optical Power Measurement. Level sensitivity; -90dBm (1300-1620nm, resolution ≥0.05nm), -85dbm (1000-1300nm, resolution ≥0.05nm), -60dBm (600 - 1000nm, resolution ≥0.05nm). Maximum input power: +20dBm. Level accuracy: ±0.4dB (1310/1550nm, input level -20dBm). Level linearity: ±0.05dB( input level: -50 to +10dBm). Wavelength sampling points (Maximum): 50,000. Optical return loss: >35dB (with angled-PC connector). Polarization dependence: ±0.05dB (1550nm). Dynamic range; Peak ±0.1nm 39dB (Resolution: 0.02nm), Peak ±0.4nm 60dB (Resolution: 0.05nm).
|