Tektronix 7B85 Horizontal time base recommended for use with 7000, 7800 and 7900 Series Mainframes to provide optimum bandwidth/ sweep-speed compatibility. Features: Delta Time Measurements with Crt Readout. Delay Time Measurements with Crt Readout. Vertical Trace Separation between Two Delayed Sweeps. 1 ns/div to 5 s/div Calibrated Time Bases. Triggering to 400 MHz. Variable Trigger Holdoff. Peak-to-Peak Auto Triggering. Sweep Rates - 5 s/div to 10 ns/div in 27 steps (1-2-5 sequence). 10X MAGNIFIER extends fastest calibrated sweep rate to 1 ns div. The uncalibrated VARIABLE is continuous to at least 2.5 times the calibrated sweep. Trace Separation Range - Funclional only in Delta Delay Time mode when alternating or chopping between time-base units. The second delayed sweep display can be vertically positioned at least 3 div below the first delayed sweep display. Delay Time Range - 0.2 or less to at least 9.0 times TIME/ DIV setting. Jitter - 0.02% of TIME/DIV setting + 0.1 ns, or less. External Trigger Input - Max input voltage is 250 V (dc + peak ac) Input R and C is 1 M-Ohm within 5% and 20 pF within 10%. The level range (excluding p-p AUTO) is at least ± 1.5 V in EXT + 1, and at least ±15 V in EXT + 10.
|