The Vitrek V63 5KV AC, 6KV DC Hipot Tester with Insulation Resistance and Ground Continuity comes loaded with standard features that help you automate the task at hand. With the built-in fully bidirectional RS232 interface, you can download and run test sequences, as well as retrieve test results and data. The PCL Interface 9-pin D-sub connector with outputs of Testing, Pass, Fail with inputs of Start, Reset. Standard features to assist in automated testing and data storage include: Built-in 160 Step Test Memory, Programmable Ramp & Dwell Times, and Programmable Output Voltages and Limits. Specifications. Voltage Range AC: 0.100KV to 5.000KV. Voltage Setting AC: 5 Volt Increments. Voltage Regulation (Line & Load) AC: 1% + 5V. Voltage Accuracy AC: 1% of reading + 10V. Current Sourcing AC: 30 to 40mA (above 500V, 180sec. max.), 0 to 29.99mA (above 500V, continuous), 0 to 10mA (below 500V, continuous). Current Limit Range AC: 0.10 to 40mA, 0.02mA/step. Voltage Range DC: 0.100KV to 6.000KV. Voltage Setting DC: 5 Volt Increments. Voltage Regulation (Line & Load) DC: 1% + 5V. Voltage Accuracy DC: 1% of reading + 10V. Current Sourcing DC: 0 to 10mA (above 500V, continuous), 0 to 2mA (below 500V, continuous). Current Limit Range DC: 0.10 to 10mA, 0.01mA/step. Current Accuracy DC: 1.5% of reading + 50µA . Insulation Resistance DC Output Voltage: 50V / 100V 500V / 1000V. Insulation Resistance Accuracy 50V / 100V: 1 to 50M-ohm: 5% of reading, 51 to 200 M-ohm: 10% of reading, 201 to 2000 M-ohm: 20% of reading. Insulation Resistance Accuracy 500V / 1000V: 1 to 500 M-ohm: 5% of reading, 501 to 2000 M-ohm: 10% of reading, 2001 to 9900 M-ohm: 20% of reading. Continuity Measurement Test Current: 100 mA. Continuity Measurement Resistance Range: 0.001 to 1.500 ohm. Continuity Measurement Resistance Accuracy: ±0.075 ohm. GPIB Option GP-60, IEEE-488 interface.
|