The GW Instek GPT-9804 AC/DC Withstanding Voltage / Insulation Resistance / Ground Bond Tester looks to meet a variety of safety standards such as IEC, EN, UL, CSA, GB, JIS and other safety-related requirements for the safety tests of electronic products and components. The GPT-9804 is a 4-in-1 model capable of performing AC withstanding, DC withstanding, insulation resistance and ground bond tests. It is equipped with a simple & clear panel layout, a high resolution dot matrix LCD display, and color LED indicators, allows operators to interpret measurement results easily and quickly. Specifications.
AC Output-Voltage Range: 0.100kV to 5.000kV. AC Output-Voltage Resolution: 2V/step. AC Output-Voltage Accuracy: ± (1% of setting +5V) [no load]. AC Maximum Rated Load: 200 VA (5kV/40mA). Output-Voltage Waveform: Sine wave. Output-Voltage Frequency: 50 Hz / 60 Hz selectable. DC Output-Voltage Range: 0.100kV to 6.000kV. DC Output-Voltage Resolution: 2V/step. DC Output-Voltage Accuracy: ± (1% of setting +5V) [no load]. DC Maximum Rated Load: 50W (5kV/10mA). ARC Detect: Yes. RAMP (Ramp-Up Time): 0.1 to 999.9s. TIMER (Test Time): OFF, 0.5s to 999.9s. IR Output-Voltage: 50V to 1000V DC. IR Output-Voltage Resolution: 50V/step. IR Output-Voltage Accuracy: ± (1% of setting +5V) [no load]. IR Resistance Measurement Range: 1 MΩ to 9500 MΩ. Ground Bond Output-Current Range: 3.00A to 30.00A AC. Ground Bond Output-Current Resolution: 0.01A. Ground Bond Resistance Measurement Range: 10 mΩ to 50.0 mΩ. RS232: Standard. USB (Device): Standard. GPIB: Optional.
|