The Fluke LT544Dlite Digital Safety Tester is designed to perform simple electrical safety tests on any type of medical equipment, whenever patient lead testing is not required. The rugged, hand-held instrument is built for quick and easy use. A simple selector knob controls the functions: Leakage Current and Chassis Resistance. Accurate test measurements are shown on the large display. Heavy-duty switches change polarity, open and close the neutral and select Chassis or Earth Leakage Current measurements. A four-wire Kelvin bridge eliminates reading errors in resistance due to cable length and contact resistance. True RMS measurements are provided for all current readings. The AAMI test load is utilized. Specifications. Ground Resistance Range: 0 – 19.99 ohms. Ground Resistance Accuracy: ± 1% R, ± 1 LSD. Ground Resistance Resolution: 10 mOhms. Leakage Current Range: 0 - 1,999 µA. Leakage Current Accuracy: DC & 25 Hz - 1kHz (±1% of reading, ± 3LSD), 1.0KHz - 100 KHz (±2.5% of reading, ± 3LSD), 100 KHz - 1 MHz (±5% of reading, ± 3LSD). Current Capacity: 15Amps at 120 VAC (1515), 10 Amps at 230 VAC (230V), 20 Amps at 230 VAC 20% duty cycle (2 minutes on, 8 minutes off) (1520/2020).
|