The GW Instek GPT- 9902A is capable of performing both AC and DC withstanding tests needed for the compliance of the safety standards such as IEC, EN, UL, CSA, GB, JIS and other safety regulations. The output voltage is automatically cut off (within 150μs) upon the detection of an abnormal output voltage or a trip of current limits during test to protect the operator from hazardous injury. It automatically discharges the DUT after each test to eliminate excessive voltage that remains on the DUT. Specifications.
AC Withstanding. Output-Voltage Range: 0.050kV to 5.000kV ac. Output-Voltage Resolution: 2V/step. Output-Voltage Accuracy: ± (1% of setting +5V) [no load]. Maximum Rated Load: 500 VA (5kV/100mA). Maximum Rated Current: 100mA (0.5kV < V ≤ 5kV), 10 mA (0.05kV ≤ V≤ 0.5kV). Output-Voltage Waveform: Sine wave. Output-Voltage Frequency: 50 Hz / 60 Hz selectable. Voltmeter Accuracy: ± (1% of reading+ 5V). Current Measurement Range: 0.001 mA to 100.0 mA. Current Best Resolution: 0.001mA / 0.01mA / 0.1mA. AC Current Measurement Accuracy: ±(1.5% of reading +30 counts) when HISET < 1.11mA, ±(1.5% of reading +3 counts) when HISET ≥ 1.11mA. ARC Detect: Yes. RAMP (Ramp-Up Time): 0.1 to 999.9s. TIMER (Test Time): OFF, 0.5s to 999.9s. Sweep Function: Yes. GND: ON/OFF.
DC Withstanding. Output-Voltage Range: 0.050kV to 6.000kV dc. Output-Voltage Resolution: 2V/step. Output-Voltage Accuracy: ± (1% of setting +5V) [no load]. Maximum Rated Load: 100W (5kV/20mA). Maximum Rated Current: 20mA (0.5kV < V ≤ 6kV), 2mA (0. 05kV ≤ V≤ 0.5kV). Voltmeter Accuracy: ± (1% of reading+ 5V). Current Measurement Range: 0.001mA to 20.0mA. Current Best Resolution: 0.001mA / 0.01mA / 0.1mA. DC Current Measurement Accuracy: ±(1.5% of reading +30 counts) when HISET < 1.11mA, ±(1.5% of reading +3 counts) when HISET ≥ 1.11mA. ARC Detect: Yes. RAMP (Ramp-Up Time): 0.1 to 999.9s. TIMER (Test Time): OFF, 0.5s to 999.9s. Sweep Function: Yes. GND: ON/OFF.
|