The Associated Research 3565D Hypot II Dielectric Withstand tester has no load setup of trip current and output voltage. The Hypot II includes rear panel remote control connections to make it easier to build into an automated system. As an option, rear panel high voltage and return terminals can also be provided. Specifications. Output Rating AC: 0-5000V, 2V/step, 12 mA. Output Rating DC: 0-6000V, 2V/step, 5 mA. Output Regulation: ± (1% of output + 5V). Voltage Setting: 0V-Max output rating, 10 volts/steps. Voltage Setting Accuracy: ± (2% of Setting + 5V). Output Frequency: DC and 50/60 Hz selectable. Waveform: Sine wave, Distortion: < 2% THD. Ripple: < 5% at 6 KV DC/5 mA (DC mode only). Dwell Time Setting: 0 and 0.2-999.9 second, 0.1 second/steps, "0" for continuous running. Ramp Time Setting: 0.2 - 999.9 seconds in 0.1 second increments, 0 ramp setting = 0.1 seconds fixed. Voltmeter (3 digits) Range: AC 0.00 - 5.00 KV, DC 0.00 - 6.00 KV. Ammeter (4 digits) Range: AC 0.10 - 12.00 mA, DC 0.02 - 5.00 mA. Ground Continuity Check: Current: DC 0.1A ± 0.01A, fixed, Max. ground resistance: 1 ohm ± 0.1 ohm, fixed. Low Current Sense: 0.00 - 12.00 mA AC, 0.00 - 5.00 mA DC. Low Current Sense Accuracy: ± (2% of setting + 0.02 mA). Options. 1-01, Rear Panel Outputs (HV, Return, Continuity). 2-03, Remote Interlock. 3-04, Arc Detection Function.
|