The Datron 4200A Autocal AC Standard calibrator features 90 day, ± 1 °C total uncertainties of ±100 ppm. AC Voltage Ranges: 1 mV to 1000V in decades. Full Scale: 2 x range except 1000V range, where max output 1100V. Resolution: 1 digit in 1,999,999 or 100 nV, whichever is greater. Frequency: Ranges: 100Hz to 1MHz in decade steps. Resolution: 1 % of range. Accuracy < ±100ppm. Sensing: Selectable remote/ local sensing on 1 V to 1000V ranges. Guarding: Selectable remote/ local guarding. Maximum Capacitive load: 1000 pF on 1V to 100V ranges, 300 pF on 1000V range. AC Current (option 30); Ranges: 100 uA to 1 A in decades. Full scale: 2 x range. Resolution: 1 digit in 1,999,999 or 100 pA, whichever is greater. Guarding: Selectable remote/ local guarding. Calibration: Autocal from front panel or via the IEEE-488 interface. OPTIONS; 10: 1000V Range. 30: Current Function. 42: Alternative Rear Output.
|