The Fluke 701 Documenting Process Calibrator is a battery-powered, hand-held instrument that lets you troubleshoot, calibrate, verify, and document the calibration of process instruments. It has an analog display for easy to read measurements when input levels are unstable. Thermocouple (TC) (E, N, J, L, K, T, U, B, R, S, or C) input/output jack and internal isothermal block with automatic reference-junction temperature compensation or manually record an external temperature reference. Specifications.
DC Voltage Measurement. Input Impedance: 5 MΩ. Common Mode Error: 0.008% f.s./(Common Mode Volt). Maximum Input Voltage: 300 V rms. Range / Resolution / Accuracy, (% of Reading +% of Full Scale (1 year)). 110 mV / 1 μV / 0.05% + 0.015%, 1.1 V / 10 μV / 0.05% + 0.005%, 11 V / 100 μV / 0.05% + 0.005%, 110 V / 1 mV / 0.05% + 0.005%, 300 V / 10 mV / 0.05% + 0.005%.
AC Voltage Measurement. Ranges : 1.1000 V, 11.000 V, 110.00 V, 300.0 V rms. Resolution: 11.000 counts in all ranges except 300 V; 3,000 counts on 300 V range. Input Impedance: 5 MΩ and <100 pF. Input Coupling: ac. Maximum Input Voltage: 300 V rms. Frequency Range / Accuracy, (% of Reading + Number of Counts (1 year)). 20 Hz to 40 Hz / 2% + 10, 40 Hz to 500 Hz / 0.5% + 5, 500 Hz to 1 kHz / 2% + 10, 1 kHz to 5 kHz / 10% + 20.
DC Current Measurement. Common Mode Error: 0.01% f.s./(Common Mode Volt). Maximum Input Voltage: 30 V dc. Range / Resolution / Accuracy, (% of Reading +% of Full Scale (1 year)). 30 mA / 1 μA / 0.025% +0.025%, 110 mA / 10 μA / 0.05% + 0.05%.
Resistance Measurement. Common Mode Error: 0.005% f.s/(Common Mode Volt). Maximum Input Voltage: 30 V dc. Range / Resolution / Accuracy, (% of Reading + ohms (1 year)). 11 Ω / 0.001 Ω / 0.1% + 0.05, 110 Ω / 0.01 Ω / 0.1% + 0.05, 1.1 kΩ / 0.1 Ω / 0.1% + 0.5, 11 kΩ / 1 Ω / 0.1% + 10.
Frequency Measurement. Amplitude range for frequency measurement: <1 kHz: 110 mV to 300V, 1 kHz to 30 kHz: 0.5 to 30V, >30 kHz: 1V to 30V. Input Impedance: 5 MΩ. Range / Accuracy, Counts (1 year). 1 Hz to 109.99 Hz / 5, 110 Hz to 1099.9 Hz / 5, 1.100 kHz to 10.999 kHz / 5, 11 kHz to 50 kHz / 5.
DC Voltage Output. Maximum Output Current: 10 mA. Loading: (0.001% f.s. + 1 μV)/ mA. Common Mode Error: 0.008% f.s/(Common Mode Volt). Maximum Input Voltage: 30 V dc. Range / Resolution / Accuracy, (% Output +% of Full Scale (1 year)). 110 mV / 1 μV / 0.015% + 0.005%, 1.1 V / 10 μV / 0.015% + 0.005%, 11 V / 100 μV / 0.015% + 0.005%.
DC Current Output. Maximum Burden Voltage: 24 V. Common Mode Error: 0.008% f.s/(Common Mode Volt). Maximum Input Voltage: 30 V dc. Range / Resolution / Accuracy, (% of Output +% of Full Scale (1 year)). 22 mA / 1 μA / 0.01% + 0.04%.
Resistance Sourcing. Maximum/Minimum Current through Source Resistance: 11 Ω Range: 3 mA dc max, 0.1 mA dc min; 110 Ω Range: 3 mA dc max, 0.1 mA dc min; 1.1 kΩ Range: 3 mA dc max, 0.01 mA dc min; 11 kΩ Range: 1 mA dc max, 0.01 mA dc min. Common Mode Error: 0.008% f.s/(Common Mode Volt). Maximum Input Voltage: 30V dc. Range / Resolution / Accuracy, (% of Output + Ohms (1 year)). 11.000 Ω / 1 mΩ / 0.02 % + 0.02, 110.00 Ω / 10 mΩ / 0.02 % + 0.04, 1.1000 kΩ / 100 mΩ / 0.05 % + 0.5, 11.000 kΩ / 1 Ω / 0.1 % + 5.
Frequency Sourcing. Waveform: Squarewave, 50% duty cycle. Amplitude: 0 to 10V pk. Amplitude Accuracy: 2 to 1099 Hz: 3% of output + 0.05 V, 1.1 to 10.9 kHz: 10% of output + 0.05 V, 11 to 50 kHz: 30% of output + 0.05 V. Maximum Input Voltage: 30V dc. Frequency Ranges / Count. 2 Hz to 109.99 Hz / 1, 110 Hz to 1099 Hz / 1, 1.1 kHz to 10.9 kHz / 1, 11 kHz to 50 kHz / 2.
|