FLUKE 701 CALIBRATION and FLUKE 701 REPAIR

 
A calibration by Custom-Cal is performed by engineers with extensive OEM experience. We have the expertise and the necessary standards to perform the FLUKE 701 Calibration, onsite calibration may be available. We specialize in quick turnaround times and we can handle expedited deliveries upon request.

Need another Calibrator?
Shop for Comparable Products on our
 used Calibrator page

 
   FLUKE 701   Description / Specification:    
FLUKE 701 Handheld Multifunction Documenting Process Calibrator

The Fluke 701 Documenting Process Calibrator is a battery-powered, hand-held instrument that lets you troubleshoot, calibrate, verify, and document the calibration of process instruments. It has an analog display for easy to read measurements when input levels are unstable. Thermocouple (TC) (E, N, J, L, K, T, U, B, R, S, or C) input/output jack and internal isothermal block with automatic reference-junction temperature compensation or manually record an external temperature reference.
Specifications.

DC Voltage Measurement.
Input Impedance: 5 MΩ.
Common Mode Error: 0.008% f.s./(Common Mode Volt).
Maximum Input Voltage: 300 V rms.
Range / Resolution / Accuracy, (% of Reading +% of Full Scale (1 year)).
  110 mV / 1 μV / 0.05% + 0.015%,
  1.1 V / 10 μV / 0.05% + 0.005%,
  11 V / 100 μV / 0.05% + 0.005%,
  110 V / 1 mV / 0.05% + 0.005%,
  300 V / 10 mV / 0.05% + 0.005%.

AC Voltage Measurement.
Ranges : 1.1000 V, 11.000 V, 110.00 V, 300.0 V rms.
Resolution: 11.000 counts in all ranges except 300 V; 3,000 counts on 300 V range.
Input Impedance: 5 MΩ and <100 pF.
Input Coupling: ac.
Maximum Input Voltage: 300 V rms.
Frequency Range / Accuracy, (% of Reading + Number of Counts (1 year)).
  20 Hz to 40 Hz / 2% + 10,
  40 Hz to 500 Hz / 0.5% + 5,
  500 Hz to 1 kHz / 2% + 10,
  1 kHz to 5 kHz / 10% + 20.

DC Current Measurement.
Common Mode Error: 0.01% f.s./(Common Mode Volt).
Maximum Input Voltage: 30 V dc.
Range / Resolution / Accuracy, (% of Reading +% of Full Scale (1 year)).
  30 mA / 1 μA / 0.025% +0.025%,
  110 mA / 10 μA / 0.05% + 0.05%.

Resistance Measurement.
Common Mode Error: 0.005% f.s/(Common Mode Volt).
Maximum Input Voltage: 30 V dc.
Range / Resolution / Accuracy, (% of Reading + ohms (1 year)).
  11 Ω / 0.001 Ω / 0.1% + 0.05,
  110 Ω / 0.01 Ω / 0.1% + 0.05,
  1.1 kΩ / 0.1 Ω / 0.1% + 0.5,
  11 kΩ / 1 Ω / 0.1% + 10.

Frequency Measurement.
Amplitude range for frequency measurement:
  <1 kHz: 110 mV to 300V,
  1 kHz to 30 kHz: 0.5 to 30V,
  >30 kHz: 1V to 30V.
Input Impedance: 5 MΩ.
Range / Accuracy, Counts (1 year).
  1 Hz to 109.99 Hz / 5,
  110 Hz to 1099.9 Hz / 5,
  1.100 kHz to 10.999 kHz / 5,
  11 kHz to 50 kHz / 5.

DC Voltage Output.
Maximum Output Current: 10 mA.
Loading: (0.001% f.s. + 1 μV)/ mA.
Common Mode Error: 0.008% f.s/(Common Mode Volt).
Maximum Input Voltage: 30 V dc.
Range / Resolution / Accuracy, (% Output +% of Full Scale (1 year)).
  110 mV / 1 μV / 0.015% + 0.005%,
  1.1 V / 10 μV / 0.015% + 0.005%,
  11 V / 100 μV / 0.015% + 0.005%.

DC Current Output.
Maximum Burden Voltage: 24 V.
Common Mode Error: 0.008% f.s/(Common Mode Volt).
Maximum Input Voltage: 30 V dc.
Range / Resolution / Accuracy, (% of Output +% of Full Scale (1 year)).
  22 mA / 1 μA / 0.01% + 0.04%.

Resistance Sourcing.
Maximum/Minimum Current through Source Resistance:
  11 Ω Range: 3 mA dc max, 0.1 mA dc min;
  110 Ω Range: 3 mA dc max, 0.1 mA dc min;
  1.1 kΩ Range: 3 mA dc max, 0.01 mA dc min;
  11 kΩ Range: 1 mA dc max, 0.01 mA dc min.
Common Mode Error: 0.008% f.s/(Common Mode Volt).
Maximum Input Voltage: 30V dc.
Range / Resolution / Accuracy, (% of Output + Ohms (1 year)).
  11.000 Ω / 1 mΩ / 0.02 % + 0.02,
  110.00 Ω / 10 mΩ / 0.02 % + 0.04,
  1.1000 kΩ / 100 mΩ / 0.05 % + 0.5,
  11.000 kΩ / 1 Ω / 0.1 % + 5.

Frequency Sourcing.
Waveform: Squarewave, 50% duty cycle.
Amplitude: 0 to 10V pk.
Amplitude Accuracy:
  2 to 1099 Hz: 3% of output + 0.05 V,
  1.1 to 10.9 kHz: 10% of output + 0.05 V,
  11 to 50 kHz: 30% of output + 0.05 V.
Maximum Input Voltage: 30V dc.
Frequency Ranges / Count.
  2 Hz to 109.99 Hz / 1,
  110 Hz to 1099 Hz / 1,
  1.1 kHz to 10.9 kHz / 1,
  11 kHz to 50 kHz / 2.



 

Standard Calibration $565.00 *
Click on Logo for More Prices
*This is a Web introductory price for one calibration of the FLUKE 701. Price does not in most cases include measurement performance data. Pricing does include NIST traceable calibration and issue of a calibration certificate and calibration label. Pricing may vary slightly due to volume and location of laboratory supporting calibration. Volume pricing may apply. On-site fees may apply depending on logistics, location and volume of work to be completed during the visit.


Related Bench Equipment Terms and Definitions. For a complete list go to our  Terms and Definitions Page.

dBm
dBm (sometimes dBmW) is an abbreviation for the power ratio in decibels (dB) of the measured power referenced to one milliwatt (mW). Zero dBm equals one milliwatt. A 3 dB increase represents roughly doubling the power, a 3 dB decrease, the power is reduced by about one half.

Millimeter-wave
Millimeter-wave is a term that generally refers to frequencies of 26.5 GHz and above. That is where signal wavelengths fall below approximately 10 mm, and thus the terminology changes from centimeters to millimeters.

Power Repeatability
Power Repeatability is the random uncertainty in reproducing the power level after changing and re-setting the power level. The power repeatability is ± half the span (in dB) between the highest and lowest actual power. Note: - The long-term power repeatability can be obtained by taken the power repeatability and power stability into account.

Stability
Stability is the extent to which a specified property, characteristic, or parameter of a substance, device, or apparatus, remains fixed with the passage of time or with varying environmental conditions.


Please contact us for your FLUKE 701 CALIBRATION and/or FLUKE 701 Handheld Multifunction Documenting Process Calibrator REPAIR

I'm interested in: 
Calibration Type (Select one):
Repair Information:
 Serial Number:

 Is there any physical damage?  
 If yes, describe:


What is the description of the problem?

How/when did the problem start?


 (*) Required Fields    For formal quote please fill in all fields.
Company:
From (Email):*
Contact Name:*
Country:*
Address 1:
Address 2:
City:
State/Territory/Province: *
Zip/Postal Code:
Phone:*

Comment: