YOKOGAWA AQ6370D REPLACEMENT / ALTERNATE

Custom-Cal has a solution to replace your YOKOGAWA AQ6370D. We specialize in quick turnaround times and we can handle expedited deliveries upon request.

The Deviser Optical Spectrum Analyzer is a Monochromatic (Grating Filter) type OSA designed to cover O,C and L optical communication bands as well as visible light starting at 600nM.  (600 ~ 1700nm).    Specification are comparable to the latest model in the Yokogawa AQ6370 series at a price that fits your budget.  Some measurement types this OSA can easily accommodate are: Optical Amplifier (EDFA) Measurement, Fabry-Perot Laser Diode (FP LD) Analysis, Distributed Feedback Laser Diode (DFB LD) Analysis, Gas measurement, Wavelength-Division Multiplexing (WDM) Channel Analysis and Notch Width Measurement.
Original
Replacement
  
YOKOGAWA AQ6370D 600 to 1700 nm Optical Spectrum Analyzer (OSA)

DEVISER AE8600 600 to 1700 nm Optical Spectrum Analyzer

YOKOGAWA AQ6370D   Description / Specification: DEVISER AE8600   Description / Specification:      
The Yokogawa AQ6370D Optical Spectrum Analyzer (OSA) can measure high power sources such as optical amplifiers and pump lasers for Raman amplifiers, and very weak optical signals as well. Measurement sensitivity can be chosen from seven categories according to test applications and measurement speed requirements. With an advanced monochromator, faster electrical circuits, and noise reduction techniques, the AQ6370D achieves fast measurement speed even when measuring a steep spectrum from DFB-LD or DWDM signals, or when measuring a low power signal from a broadband light source.
Specifications.
Wavelength range: 600 to 1700 nm.
Span: 0.1 nm to 1100 nm (Full span), and 0 nm.
Wavelength accuracy: ±0.02 nm (1520 to 1620 nm), ±0.04 nm (1450 to 1520 nm), ±0.10 nm (Full range).
Wavelength linearity: ±0.01 nm (1520 to 1580 nm), ±0.02 nm (1450 to 1520 nm, 1580 to 1620 nm).
Wavelength repeatability: ±0.005 nm (1 min.).
Wavelength resolution setting: 0.02, 0.05, 0.1, 0.2, 0.5, 1 and 2 nm.
Wavelength resolution bandwidth accuracy: ±5% (1450 to 1620 nm, Resolution setting: ≥ 0.1 nm, after performing the Resolution Calibration function, at the wavelength of resolution calibration).
Min. sampling resolution: 0.001 nm.
Number of sampling: 101 to 50001, AUTO.
Level sensitivity setting: NORM_HOLD, NORM_AUTO, NORMAL, MID, HIGH1, HIGH2 and HIGH3.
High dynamic mode: SWITCH (Sensitivity: MID, HIGH1-3).
Level sensitivity: −90 dBm (1300 to 1620 nm), −85 dBm (1000 to 1300 nm), −60 dBm (600 to 1000 nm) (Sensitivity: HIGH3).
Maximum input power: +20 dBm (Per channel, full range).
Maximum safe input power: +25 dBm (Total input power).
Level accuracy: ±0.4 dB (1310/1550 nm, Input level: −20 dBm, Sensitivity: MID, HIGH1-3).
Level linearity: ±0.05 dB (Input level: −50 to +10 dBm, Sensitivity: HIGH1-3).
Level flatness: ±0.1 dB (1520 to 1580 nm), ±0.2 dB (1450 to 1520 nm, 1580 to 1620 nm).
Polarization dependence: ±0.05 dB (1550/1600 nm), ±0.08 dB (1310 nm).
Stray-light suppression ratio: 73 dB.
Optical return loss: Typ. 35 dB (with angled-PC connector).
Applicable fiber: SM (9.5/125 μm), GI (50/125 μm, 62.5/125 μm).

Suffix codes.
-10 Standard model.
-20 High performance model.
-L0 Without light source.
-L1 Wavelength reference source.


The Deviser E8600 is a high-precision diffraction-grating, high-resolution optical spectrometer with wavelength range of 600 nm to 1700 nm. The 10.1” LCD touchscreen and concise graphical user interface of AE8600 offer the easiest way to handle optical spectrum analysis. It provides a wide selection of test methodology, including laser spectrum scans (DPB, FP), WDM system testing, EDFA system testing, transmittance and drift testing, which are essential for in-field and factory applications. The AE8600 offers exceptional stability and reliability, high-speed spectral sweeping, and multiple ways to export and analyze measurement data. It’s the ideal instrument for fast and precise optical spectral testing to satisfy long-term investment with the best cost performance value.
Specifications.

Optical Spectrum Measurement.
Wavelength range: 600 to 1700nm.
Wavelength resolution bandwidth: 0.02 to 2nm.
Wavelength resolution setting: 0.02nm, 0.06nm, 0.1nm, 0.2nm, 0.5nm, 1nm, 2nm.
Wavelength accuracy;
  1520 to 1620 nm ±0.02 nm,
  1450 to 1520 nm ±0.04 nm,
  Entire wavelength range ±0.1 nm.
Wavelength repeatability: ±0.005 nm (1 minute).
Wavelength linearity;
  ±0.01 nm (1520 to 1580 nm),
  ±0.02 nm (1450 to 1520 nm, 1580 to 1620 nm).
Minimum sampling resolution: 0.001nm.
Applicable fiber: SM(9.5/125μm), MMF(50/125μm, 62.5/125μm).

Optical Power Measurement.
Level sensitivity;
  -90dBm (1300-1620nm, resolution ≥0.05nm),
  -85dbm (1000-1300nm, resolution ≥0.05nm),
  -60dBm (600 - 1000nm, resolution ≥0.05nm).
Maximum input power: +20dBm.
Level accuracy: ±0.4dB (1310/1550nm, input level -20dBm).
Level linearity: ±0.05dB( input level: -50 to +10dBm).
Wavelength sampling points (Maximum): 50,000.
Optical return loss: >35dB (with angled-PC connector).
Polarization dependence: ±0.05dB (1550nm).
Dynamic range;
  Peak ±0.1nm 39dB (Resolution: 0.02nm),
  Peak ±0.4nm 60dB (Resolution: 0.05nm).