Custom-Cal has a high success rate in repairing Agilent's 8560 Series of Spectrum Analyzers. We specialize in quick turnaround times and we can handle expedited deliveries upon request.
The HP 8560 EC-series portable spectrum analyzers including the ( 8560EC, 8561EC, 8562EC, 8563EC, 8564EC, 8565EC )combine outstanding phase noise, sensitivity, 1 Hz resolution bandwidths, and wide dynamic range in a MIL-rugged package built to withstand harsh environmental conditions. They offer a complete solution for ACP testing of burst carrier signals using digital modulation such as is used in NADC-TDMA, GSM, DECT, CT2-CAI, PDC and PHP system. With digitally implemented resolution bandwidths of 1, 3, 10, 30, and 100 Hz they sweep from 3 to 600 times faster than is possible with comparable analog filters. A narrow 5:1 shape factor allows you to view close-in, low-level signals easily. Digital bandwidths also provide the spectrum analyzer with a full 100 dB onscreen calibrated display. Digitized fast time-domain (zero span) sweeps use markers, trace math, trace storage and get hardcopy output, for measurements such as rise/fall times, pulse widths and time between events.

Errors and Options for Spectrum Analyzer.
  Error Messages:
ERR 100
  NO PWRON Power-on state is invalid; default state is loaded.
ERR 101
  NO STATE State to be RECALLed not valid or not SAVEd.
ERR 102
  # ARGMTS Command does not have enough arguments.
ERR 103
  # ARGMTS Command does not have enough arguments.
ERR 104
  # ARGMTS Command does not have enough arguments.
ERR 105
  # ARGMTS Command does not have enough arguments.
ERR 106
  ABORTED! Current operation is aborted; GPIB parser reset.
ERR 107
  HELLO ?? No GPIB listener is present.
ERR 108
  TIME OUT Analyzer timed out when acting as controller.
ERR 109
  CtrlFail Analyzer unable to take control of the bus.
ERR 110
  NOT CTRL Analyzer is not system controller.
ERR 111
  # ARGMTS Command does not have enough arguments.
ERR 112
  ??CMD?? Unrecognized command.
ERR 113
  FREQ NO! Command cannot have frequency units.
ERR 114
  TIME NO! Command cannot have time units.
ERR 115
  AMPL NO! Command cannot have amplitude units.
ERR 116
  ?UNITS?? Unrecognizable units.
ERR 117
  NOP NUM Command cannot have numeric units.
ERR 118
  NOP EP Enable parameter cannot be used.
ERR 119
  NOP UPDN UP/DN are not valid arguments for this command.
ERR 120
  NOP ONOF ON/OFF are not valid arguments for this command.
ERR 121
  NOP ARG AUTO/MAN are not valid arguments for this command.
ERR 122
  NOP TRC Trace registers are not valid for this command.
ERR 123
  NOP ABLK A-block format not valid here.
ERR 124
  NOP IBLK I-block format not valid here.
ERR 125
  NOP STRNG Strings are not valid for this command.
ERR 126
  NO ? This command cannot be queried.
ERR 127
  BAD DTMD Not a valid peak detector mode.
ERR 128
  PK WHAT? Not a valid peak search parameter.
ERR 129
  PRE TERM Premature A-block termination.
ERR 130
  BAD TDF Arguments are only for TDF command.
ERR 131
  ?? AM/FM AM/FM are not valid arguments for this command.
ERR 132
  !FAV/RMP FAV/RAMP are not valid arguments for this command.
ERR 133
  !INT/EXT INT/EXT are not valid arguments for this command.
ERR 134
  ??? ZERO ZERO is not a valid argument for this command.
ERR 135
  ??? CURR CURR is not a valid argument for this command.
ERR 136
  ??? FULL FULL is not a valid argument for this command.
ERR 137
  ??? LAST LAST is not a valid argument for this command.
ERR 138
  !GRT/DSP GRT/DSP are not valid arguments for this command.
ERR 139
  PLOTONLY Argument can only be used with PLOT command.
ERR 140
  ?? PWRON PWRON is not a valid argument for this command.
ERR 141
  BAD ARG Argument can only be used with FDIAG command.
ERR 142
  BAD ARG Query expected for FDIAG command.
ERR 143
  NO PRESL No preselector hardware to use command with.
ERR 144
  COUPL?? Invalid COUPLING argument, expected AC or DC.
ERR 200
  SYSTEM ADC Driver/ADC hardware/firmware interaction; check for other errors.
ERR 201
  SYSTEM ADC Controller/ADC hardware/firmware interaction; check for other errors.
ERR 202
  FADC CAL Linear offset search failed. This error applies applies only to EC-series analyzers and E-series analyzers with fast ADC (Option 007)
ERR 203
  FADC CAL Log offset search failed. This error applies applies only to EC-series analyzers and E-series analyzers with fast ADC (Option 007)
ERR 204
  FADC CAL Log 2´ expand offset search failed. This error applies applies only to EC-series analyzers and E-series analyzers with fast ADC (Option 007)
ERR 205
  FADC CAL Linear slope calibration is out of range. This error applies applies only to EC-series analyzers and E-series analyzers with fast ADC (Option 007)
ERR 206
  FADC CAL Log slope calibration is out of range. This error applies applies only to EC-series analyzers and E-series analyzers with fast ADC (Option 007)
ERR 207
  FADC CAL Log 2´ expand slope calibration is out of range. This error applies applies only to EC-series analyzers and E-series analyzers with fast ADC (Option 007)
ERR 250
  OUTOF RG ADC input is outside of ADC range.
ERR 251
  NO IRQ Microprocessor not receiving interrupt from ADC.
ERR 300
  YTO UNLK YTO (1ST LO) phase-locked loop (PLL) is unlocked.
ERR 301
  YTO UNLK YTO PLL is unlocked.
ERR 302
  OFF UNLK Offset roller oscillator PLL is unlocked.
ERR 303
  XFR UNLK Transfer roller oscillator PLL is unlocked.
ERR 304
  ROL UNLK Main roller oscillator PLL is unlocked.
ERR 305
  FREQ ACC Coarse adjust DAC cannot bring MAINSENSE close to zero.
ERR 306
  FREQ ACC Fine adjust DAC cannot bring MAINSENSE close to zero.
ERR 307
  FREQ ACC Transfer oscillator pretune DAC out of range.
ERR 308
  FREQ ACC Offset oscillator pretune DAC not within prescribed limits at low frequency.
ERR 309
  FREQ ACC Offset oscillator pretune DAC not within prescribed limits at high frequency.
ERR 310
  FREQ ACC Main roller pretune DAC set to 255 before MAINSENSE changes to negative polarity.
ERR 311
  FREQ ACC Main roller pretune DAC set to 255 before MAINSENSE changes to negative polarity.
ERR 312
  FREQ ACC Fine adjust DAC cannot bring MAINSENSE close to zero.
ERR 313
  FREQ ACC The combination of sampler oscillator and roller oscillator frequencies do not correspond to the required YTO start frequency.
ERR 314
  FREQ ACC Span calibration problems. An unlocked main roller loop or lack of a sweep ramp.
ERR 315
  FREQ ACC Span calibration problems. Roller span attenuator DAC is out of range.
ERR 316
  FREQ ACC Sensitivity of the main roller DAC is zero.
ERR 317
  FREQ ACC Main coil coarse DAC at limits.
ERR 318
  FREQ ACC Main coil fine DAC at limits.
ERR 319
  WARN COA The YTO coarse tune DAC is near its limits.
ERR 320
  WARN FIN The YTO fine tune DAC is near its limits.
ERR 321
  FREQ ACC Main roller tuning sensitivity is not greater than zero.
ERR 322
  FREQ ACC Main roller pretune DAC value set greater than 255.
ERR 324
  FREQ ACC Coarse adjust DAC cannot bring MAINSENSE close to zero.
ERR 325
  FREQ ACC Fine adjust DAC cannot bring MAINSENSE close to zero.
ERR 326
  FREQ ACC Fine adjust DAC near the end of range.
ERR 327
  OFF UNLK Offset roller oscillator PLL is unlocked.
ERR 328
  FREQ ACC Roller fine adjust DAC sensitivity less than or equal to zero.
ERR 329
  FREQ ACC Roller coarse adjust DAC sensitivity less than or equal to zero.
ERR 331
  FREQ ACC Invalid YTO frequency.
ERR 333
  600 UNLK 600 MHz reference oscillator PLL is unlocked.
ERR 334
  LO AMPL 1st LO distribution amplifier is unleveled. The front panel LO OUTPUT must be terminated in a 50W load.
ERR 335
  SMP UNLK Sampling oscillator PLL is unlocked.
ERR 336
  10 MHz Ref Cal oscillator failed to lock when going to internal 10 MHz reference.
ERR 337
  FN UNLK Fractional N circuitry is unable to lock.
ERR 338
  LOMA AGC The LOMA (local oscillator multiplier amplifier) main loop AGC is unleveled.
ERR 339
  SBTX AGC The LOMA SBTX (switched barium-tuned mixer) loop AGC is unleveled.
ERR 340
  SAMP AGC The LOMA (local oscillator multiplier amplifier) sampler loop AGC is unleveled.
ERR 351
  SETL FLD The YTO error voltage is not settling.
ERR 352
  TWID FLD Unable to bring error voltage DACs to quiescent point.
ERR 353
  SRCH FLD No acceptable YTO DAC value found.
ERR 354
  LK ITERS Cannot lock. Lock iteration routine terminated.
ERR 355
  SMP CAL Sampler unlock condition during calibration routine. This error remains until a successful recalibration is performed.
ERR 356
  SPAC CAL Sweep data problem finding "bucket 1" of the span accuracy calibration sweep.
ERR 357
  SPAC CAL Cannot find the "x" intersection for "bucket 1" of the span accuracy calibration sweep.
ERR 358
  SPAC CAL Cannot find "bucket 2" of the span accuracy calibration sweep.
ERR 359
  SPAC CAL Cannot find the "x" intersection for "bucket 2" of the span accuracy calibration sweep.
ERR 360
  SPAC CAL The start bucket correction is out of range.
ERR 361
  SPAC CAL The percent of span correction is out of range.
ERR 400
  AMPL<300 Unable to adjust less than 300 Hz RES BW amplitude.
ERR 401
  AMPL 300 Unable to adjust amplitude of 300 Hz RES BW.
ERR 402
  AMPL 1K Unable to adjust amplitude of 1 kHz RES BW.
ERR 403
  AMPL 3K Unable to adjust amplitude of 3 kHz RES BW.
ERR 404
  AMPL 10K Unable to adjust amplitude of 10 kHz RES BW.
ERR 405
  RBW 10K Unable to adjust 10 kHz RES BW pole 1.
ERR 406
  RBW 10K Unable to adjust 10 kHz RES BW pole 2.
ERR 407
  RBW 10K Unable to adjust 10 kHz RES BW pole 3.
ERR 408
  RBW 10K Unable to adjust 10 kHz RES BW pole 4.
ERR 409
  RBW 10K Unable to adjust 10 kHz RES BW pole 1.
ERR 410
  RBW 10K Unable to adjust 10 kHz RES BW pole 2.
ERR 411
  RBW 10K Unable to adjust 10 kHz RES BW pole 3.
ERR 412
  RBW 10K Unable to adjust 10 kHz RES BW pole 4.
ERR 413
  RBW 10K Unable to adjust 10 kHz RES BW pole 1.
ERR 414
  RBW 10K Unable to adjust 10 kHz RES BW pole 2.
ERR 415
  RBW 10K Unable to adjust 10 kHz RES BW pole 3.
ERR 416
  RBW 10K Unable to adjust 10 kHz RES BW pole 4.
ERR 417
  RBW 3K Unable to adjust 3 kHz RES BW pole 1.
ERR 418
  RBW 3K Unable to adjust 3 kHz RES BW pole 2.
ERR 419
  RBW 3K Unable to adjust 3 kHz RES BW pole 3.
ERR 420
  RBW 3K Unable to adjust 3 kHz RES BW pole 4.
ERR 421
  RBW 10K Unable to adjust 10 kHz RES BW pole 1.
ERR 422
  RBW 10K Unable to adjust 10 kHz RES BW pole 2.
ERR 423
  RBW 10K Unable to adjust 10 kHz RES BW pole 3.
ERR 424
  RBW 10K Unable to adjust 10 kHz RES BW pole 4.
ERR 425
  RBW 3K Unable to adjust 3 kHz RES BW pole 1.
ERR 426
  RBW 3K Unable to adjust 3 kHz RES BW pole 2.
ERR 427
  RBW 3K Unable to adjust 3 kHz RES BW pole 3.
ERR 428
  RBW 3K Unable to adjust 3 kHz RES BW pole 4.
ERR 429
  RBW <300 Unable to adjust less than 300 Hz RES BW.
ERR 430
  RBW 300 Unable to adjust 300 Hz RES BW.
ERR 431
  RBW 1K Unable to adjust 1 kHz RES BW.
ERR 432
  RBW 3K Unable to adjust 3 kHz RES BW.
ERR 433
  RBW 10K Unable to adjust 10 kHz RES BW.
ERR 434
  RBW 300 Unable to adjust 300 Hz RES BW pole 1.
ERR 435
  RBW 300 Unable to adjust 300 Hz RES BW pole 2.
ERR 436
  RBW 300 Unable to adjust 300 Hz RES BW pole 3.
ERR 437
  RBW 300 Unable to adjust 300 Hz RES BW pole 4.
ERR 438
  RBW 1K Unable to adjust 1 kHz RES BW pole 1.
ERR 439
  RBW 1K Unable to adjust 1 kHz RES BW pole 2.
ERR 440
  RBW 1K Unable to adjust 1 kHz RES BW pole 3.
ERR 441
  RBW 1K Unable to adjust 1 kHz RES BW pole 4.
ERR 442
  RBW 3K Unable to adjust 3 kHz RES BW pole 1.
ERR 443
  RBW 3K Unable to adjust 3 kHz RES BW pole 2.
ERR 444
  RBW 3K Unable to adjust 3 kHz RES BW pole 3.
ERR 445
  RBW 3K Unable to adjust 3 kHz RES BW pole 4.
ERR 446
  RBW 10K Unable to adjust 10 kHz RES BW pole 1.
ERR 447
  RBW 10K Unable to adjust 10 kHz RES BW pole 2.
ERR 448
  RBW 10K Unable to adjust 10 kHz RES BW pole 3.
ERR 449
  RBW 10K Unable to adjust 10 kHz RES BW pole 4.
ERR 450
  IF SYSTM IF hardware failure. Check other error messages.
ERR 451
  IF SYSTM IF hardware failure. Check other error messages.
ERR 452
  IF SYSTM IF hardware failure. Check other error messages.
ERR 454
  AMPL Unable to adjust step gain amplifiers.
ERR 455
  AMPL Unable to adjust step gain amplifiers.
ERR 456
  AMPL Unable to adjust step gain amplifiers.
ERR 457
  AMPL Unable to adjust step gain amplifiers.
ERR 458
  AMPL Unable to adjust step gain amplifiers.
ERR 459
  AMPL Unable to adjust step gain amplifiers.
ERR 460
  AMPL Unable to adjust step gain amplifiers.
ERR 461
  AMPL Unable to adjust step gain amplifiers.
ERR 462
  AMPL Unable to adjust step gain amplifiers.
ERR 463
  AMPL Unable to adjust step gain amplifiers.
ERR 464
  AMPL Unable to adjust step gain amplifiers.
ERR 465
  AMPL Unable to adjust step gain amplifiers.
ERR 466
  LIN AMPL Unable to adjust linear amplitude scale.
ERR 467
  LOG AMPL Unable to adjust log amplitude scale.
ERR 468
  LOG AMPL Unable to adjust log amplitude scale.
ERR 469
  LOG AMPL Unable to adjust log amplitude scale.
ERR 470
  LOG AMPL Unable to adjust log amplitude scale.
ERR 471
  RBW 30K Unable to adjust 30 kHz RES BW.
ERR 472
  RBW 100K Unable to adjust 100 kHz RES BW.
ERR 473
  RBW 300K Unable to adjust 300 kHz RES BW.
ERR 474
  RBW 1M Unable to adjust 1 MHz RES BW.
ERR 475
  RBW 30K Unable to adjust 30 kHz RES BW.
ERR 476
  RBW 100K Unable to adjust 100 kHz RES BW.
ERR 477
  RBW 300K Unable to adjust 300 kHz RES BW.
ERR 478
  RBW 1M Unable to adjust 1 MHz RES BW.
ERR 483
  RBW 10K Unable to adjust 10 kHz RES BW.
ERR 484
  RBW 3K Unable to adjust 3 kHz RES BW.
ERR 485
  RBW 1K Unable to adjust 1 kHz RES BW.
ERR 486
  RBW 300 Unable to adjust 300 Hz RES BW.
ERR 487
  RBW 100 Unable to adjust 100 Hz RES BW.
ERR 488
  RBW 100 Unable to adjust 100 Hz RES BW.
ERR 489
  RBW 100 Unable to adjust 100 Hz RES BW.
ERR 490
  RBW 100 Unable to adjust 100 Hz RES BW.
ERR 491
  RBW <300 Unable to adjust less than 300 Hz RES BW.
ERR 492
  RBW 300 Unable to adjust 300 Hz RES BW.
ERR 493
  RBW 1K Unable to adjust 1 kHz RES BW.
ERR 494
  RBW 3K Unable to adjust 3 kHz RES BW.
ERR 495
  RBW 10K Unable to adjust 10 kHz RES BW.
ERR 496
  RBW 100 Unable to adjust 100 Hz RES BW.
ERR 497
  RBW 100 Unable to adjust 100 Hz RES BW.
ERR 498
  RBW 100 Unable to adjust 100 Hz RES BW.
ERR 499
  CAL UNLK Cal oscillator is unlocked.
ERR 500
  AMPL 30K Unable to adjust amplitude of 30 kHz RES BW.
ERR 501
  AMPL .1M Unable to adjust amplitude of 100 kHz RES BW.
ERR 502
  AMPL .3M Unable to adjust amplitude of 300 kHz RES BW.
ERR 503
  AMPL 1M Unable to adjust amplitude of 1 MHz RES BW.
ERR 504
  AMPL 30K Unable to adjust amplitude of 30 kHz RES BW.
ERR 505
  AMPL .1M Unable to adjust amplitude of 100 kHz RES BW.
ERR 506
  AMPL .3M Unable to adjust amplitude of 300 kHz RES BW.
ERR 507
  AMPL 1M Unable to adjust amplitude of 1 MHz RES BW.
ERR 508
  AMPL 30K Insufficient gain during LC BW Cal of 30 kHz RES BW.
ERR 509
  AMPL .1M Insufficient gain during LC BW Cal of 100 kHz RES BW.
ERR 510
  AMPL .3M Insufficient gain during LC BW Cal of 300 kHz RES BW.
ERR 511
  AMPL 1M Insufficient gain during LC BW Cal of 1 MHz RES BW.
ERR 512
  RBW <300 Insufficient gain during crystal BW Cal of less than 300 Hz RES BW.
ERR 513
  RBW 300 Insufficient gain during crystal BW Cal of 300 Hz RES BW.
ERR 514
  RBW 1K Insufficient gain during crystal BW Cal of 1 kHz RES BW.
ERR 515
  RBW 3K Insufficient gain during crystal BW Cal of 3 kHz RES BW.
ERR 516
  RBW 10K Insufficient gain during crystal BW Cal of 10 kHz RES BW.
ERR 517
  RBW 100 Unable to adjust 100 Hz RES BW.
ERR 518
  RBW 300 Unable to adjust 300 Hz RES BW.
ERR 519
  RBW 1K Unable to adjust 1 kHz RES BW.
ERR 520
  RBW 3K Unable to adjust 3 kHz RES BW.
ERR 521
  RBW 10K Unable to adjust 10 kHz RES BW.
ERR 522
  RBW 10K Unable to adjust 10 kHz RES BW symmetry of pole 1.
ERR 523
  RBW 10K Unable to adjust 10 kHz RES BW symmetry of pole 2.
ERR 524
  RBW 10K Unable to adjust 10 kHz RES BW symmetry of pole 3.
ERR 525
  RBW 10K Unable to adjust 10 kHz RES BW symmetry of pole 4.
ERR 526
  RBW <300 Unable to adjust less than 300 Hz RES BWs. Timeout during data sampling.
ERR 527
  RBW <300 Step gain correction failed for less than 300 Hz RES BW.
ERR 528
  RBW <300 Unable to adjust less than 300 Hz RES BWs. DC level at ADC cannot be calibrated.
ERR 529
  RBW <300 Unable to adjust less than 300 Hz RES BWs. Demod data for calibration is bad.
ERR 530
  RBW <300 Unable to adjust less than 300 Hz RES BWs. Narrow BW VCXO Calibration failed.
ERR 531
  RBW <300 Flatness correction data not acceptable for less than 300 Hz RES BWs.
ERR 532
  RBW <300 Absolute gain data for less than 300 Hz RES BWs not acceptable.
ERR 533
  RBW <300 Unable to adjust less than 300 Hz RES BWs. Timeout during data sampling narrow BW chunk.
ERR 534
  RBW <300 Unable to adjust less than 300 Hz RES BWs. Frequency count of CAL OSC using IF downconverter failed.
ERR 535
  RBW <300 Unable to adjust less than 300 Hz RES BWs. Inadequate FM demod range for 500Hz IF filter measurement.
ERR 536
  RBW <300 Unable to adjust less than 300 Hz RES BWs. Unable to autorange the chirp signal.
ERR 537
  RBW <300 Unable to adjust less than 300 Hz RES BWs. CW CAL OSC not autoranging.
ERR 538
  RBW <300 Unable to adjust less than 300 Hz RES BWs. Noisy shape of 500 Hz IF filter.
ERR 539
  RBW <300 Unable to adjust less than 300 Hz RES BWs. CW CAL OSC not autoranging.
ERR 540
  RBW <300 Unable to adjust less than 300 Hz RES BWs. No SIGLOCK on CW CAL OSC during pretune.
ERR 550
  IDCALOSC CAL Oscillator ID. Indicates incompatible hardware. Cal oscillator not expected.
ERR 551
  ID LOGBD LOG Board ID. Indicates incompatible hardware. Log board not expected.
ERR 552-562
  LOG AMPL Unable to adjust amplitude of log scale.
ERR 563
  LOG AMPL Unable to adjust amplitude in log scale. Third Step Gain range problem.
ERR 564-570
  LOG AMPL Unable to adjust amplitude in log scale.
ERR 571
  AMPL Unable to adjust step gain amplifiers.
ERR 572
  AMPL 1M Unable to adjust amplitude of 1 MHz RES BW.
ERR 573-577
  LOG AMPL Unable to adjust amplitude in log scale.
ERR 578
  LOG AMPL Limiter calibration error from DC logger calibration.
ERR 579
  LOG AMPL Attenuator CAL level error from DC logger calibration.
ERR 580
  LOG AMPL Calibration level error from DC logger fidelity calibration.
ERR 581
  AMPL Unable to adjust 100 kHz resolution bandwidth and resolution bandwidths less than or equal to 10 kHz.
ERR 582
  AMPL Unable to adjust 100 kHz resolution bandwidth and resolution bandwidths less than or equal to 10 kHz. Bad CALOSC calibration of sweep rate.
ERR 583
  RBW 30K Unable to adjust 30 kHz RES BW.
ERR 584
  RBW 100K Unable to adjust 100 kHz RES BW.
ERR 585
  RBW 300K Unable to adjust 300 kHz RES BW.
ERR 586
  RBW 1M Unable to adjust 1 MHz RES BW.
ERR 587
  RBW 30K Unable to adjust 30 kHz RES BW.
ERR 588
  RBW 100K Unable to adjust 100 kHz RES BW.
ERR 589
  RBW 300K Unable to adjust 300 kHz RES BW.
ERR 590
  RBW 1M Unable to adjust 1 MHz RES BW.
ERR 591
  LOG AMPL Unable to adjust amplitude in log scale.
ERR 592
  LOG AMPL Unable to adjust amplitude in log scale.
ERR 593
  LOG TUNE LIMIT Cal tune error from DC logger calibration.
ERR 594
  LOG OFST Attenuator calibration offset error from DC logger calibration.
ERR 595
  LOG ATTN Attenuator calibration absolute error from DC logger calibration.
ERR 596
  LOG FID Fidelity error from DC logger calibration.
ERR 597
  LOG OFST Fidelity offset error from DC logger calibration.
ERR 598
  LOG OFST Fidelity offset unstable from DC logger calibration.
ERR 599
  LOG GAIN Fidelity gain error from DC logger calibration.
ERR 600
  SYSTEM Hardware/firmware interaction; check other errors.
ERR 601
  SYSTEM Hardware/firmware interaction; check other errors.
ERR 650
  OUTOF RG ADC input is outside of the ADC range.
ERR 651
  NO IRQ Microprocessor is not receiving interrupt from ADC.
ERR 700
  EEROM Checksum error of EEROM A2U501.
ERR 701
  AMPL CAL Checksum error of frequency response correction data.
ERR 702
  ELAP TIM Checksum error of elapsed time data.
ERR 703
  AMPL CAL Checksum error of frequency response correction data.
ERR 704
  PRESELCT Checksum error of customer preselector peak data; external preselector data recalled in internal mode; or internal preselector data recalled in external mode.
ERR 705
  ROM U306 Checksum error of program ROM A2U306.
ERR 706
  ROM U307 Checksum error of program ROM A2U307.
ERR 707
  ROM U308 Checksum error of program ROM A2U308.
ERR 708
  ROM U309 Checksum error of program ROM A2U309.
ERR 709
  ROM U310 Checksum error of program ROM A2U310.
ERR 710
  ROM U311 Checksum error of program ROM A2U311.
ERR 711
  RAM U303 Checksum error of system RAM A2U303.
ERR 712
  RAM U302 Checksum error of system RAM A2U302.
ERR 713
  RAM U301 Checksum error of system RAM A2U301.
ERR 714
  RAM U300 Checksum error of system RAM A2U300.
ERR 715
  RAM U305 Checksum error of system RAM A2U305.
ERR 716
  RAM U304 Checksum error of system RAM A2U304.
ERR 717
  BAD uP!! Microprocessor is not fully operational.
ERR 718
  BATTERY? Nonvolatile RAM is not working; check battery.
ERR 719
  MODEL #? Cannot identify the model number.
ERR 720
  OPTION #? Cannot identify an option number.
ERR 721
  AMPC RAM Checksum error of the ampcor function correction data.
ERR 750
  SYSTEM Hardware/firmware interaction, zero divide; check other errors.
ERR 751
  SYSTEM Hardware/firmware interaction, floating overflow; check other errors.
ERR 752
  SYSTEM Hardware/firmware interaction, floating underflow; check other errors.
ERR 753
  SYSTEM Hardware/firmware interaction, LOG error; check other errors.
ERR 754
  SYSTEM Hardware/firmware interaction, Integer overflow; check other errors.
ERR 755
  SYSTEM Hardware/firmware interaction, squareroot error; check other errors.
ERR 756
  SYSTEM Hardware/firmware interaction, Triple overflow; check other errors.
ERR 757
  SYSTEM Hardware/firmware interaction, BCDL overflow; check other errors.
ERR 758
  SYSTEM Unknown system error.
ERR 759
  SYSTEM Hardware/firmware interaction. Code invoked for wrong instrument.
ERR 760
  NO FADC The power-on read of the fast analog to digital converter hardware is failing. This error applies applies only to EC-series analyzers and E-series analyzers with fast ADC (Option 007)
ERR 900
  TG UNLVL Tracking generator output is unleveled.
ERR 901
  TGFrqLmt Tracking generator output unleveled because START FREQ is set below tracking generator frequency limit (300 kHz).
ERR 902
  BAD NORM A normalization error will occur if the current spectrum analyzer state is not the same as the state stored by the last execution of the STORETHRU or STORESHORT command.
ERR 903
  A > DLMT Unnormalized trace A is off screen with trace math or normalization on.
ERR 904
  B > DLMT Calibration trace (trace B) is off screen with trace math or normalization on.
ERR 905
  EXT REF Unable to lock cal oscillator when set to external reference. Check that external reference is within tolerance.
ERR 906
  OVENCOLD The oven controlled crystal oscillator (OCXO) oven is cold.
ERR 907
  DO IFCAL Unit is still performing IF calibration, or is in need of IF calibration which were not done due to an OVENCOLD condition.
ERR 908
  BW>>SPCG Channel bandwidth is too wide, compared to the channel spacing, for a meaningful computation.
ERR 909
  SPAN<ACP The frequency span is too small to obtain a valid ACP measurement.
ERR 910
  SPAN>ACP The frequency span is too wide, compared to the channel bandwidth, to obtain a valid ACP measurement.
ERR 911
  ACP STATE The adjacent channel power measurement has been compromised (invalid measurement parameters).
ERR 919
  SP <CHBW The channel spacing is too narrow, compared to the channel bandwidth, to obtain a valid channel power bandwidth measurement.
ERR 920
  RBW>CHBW The resolution bandwidth is too wide, compared to the channel bandwidth, to obtain a valid channel power bandwidth measurement. The resolution bandwidth should be much less than the channel bandwidth (<0.1´channel BW).
ERR 921
  ­AMPCOR­ Measurement data which would normally be displayed above the top graticule, and therefore has unspecified accuracy, has been corrected by ampcor the amplitude correction function (ampcor) to appear between the top and bottom graticule.
ERR 922
  ¯AMPCOR¯ Measurement data which would normally be displayed below the bottom graticule, and therefore has unspecified accuracy, has been corrected by ampcor the amplitude correction function (ampcor) to appear between the top and bottom graticule.

Instrument Options:
001 Second IF Output (310.7 MHz)
002 Built-In Tracking Generator
005 Alternate Sweep Out (cannot be used with Option 002)
008 Signal Identification
103 Delete Precision Frequency Reference, and 1 Hz and 3 Hz RBWs
104 Delete Mass Memory Module
   
Description of Model 
Portable Spectrum Analyzer
• Frequency range: 30 Hz to 2.9 GHz
• Frequency accuracy: ± 106 Hz (CF = 1 GHz)
• Frequency Span Range: 0 Hz, 100 Hz to maximum frequency
• Frequency Readout Accuracy Span >2 MHz x N: ± (freq. readout x freq. ref. accuracy +5% x span +15% x RBW +10 Hz)
• Frequency Readout Accuracy Span ≤2 MHz x N: ± (freq. readout x freq. ref. accuracy +1% x span +15% x RBW +10 Hz)
• Sweep Time Range: Span = 0 Hz: 50 μs to 6,000 s; Span ≥100 Hz: 50 ms to 100 ks
• Resolution bandwidth range: 1 Hz to 2 MHz
• Resolution bandwidth Accuracy: 1 Hz to 300 kHz: ±10%; 1 MHz: ±25%; 2 MHz: +50%, –25%
• Video Bandwidth Range: 1 Hz to 3 MHz in a 1, 3, 10 sequence
• Amplitude Range: Displayed average noise level to +30 dBm
• Average noise level (narrowest RBW): -145 dBm
• Optimum dynamic range 2nd/3rd order: 88 dB / 103 dB
• Amplitude accuracy: ± 1.85dB

I'm interested in: 
Calibration Type (Select one):
Repair Information:
 Serial Number:

 Is there any physical damage?  
 If yes, describe:


What is the description of the problem?

How/when did the problem start?


 (*) Required Fields    For formal quote please fill in all fields.
Company:
From (Email):*
Contact Name:*
Country:*
Address 1:
Address 2:
City:
State/Territory/Province: *
Zip/Postal Code:
Phone:*

Comment: