ERR 100 |
|
NO PWRON Power-on state is invalid; default state is loaded. |
ERR 101 |
|
NO STATE State to be RECALLed not valid or not SAVEd. |
ERR 102 |
|
# ARGMTS Command does not have enough arguments. |
ERR 103 |
|
# ARGMTS Command does not have enough arguments. |
ERR 104 |
|
# ARGMTS Command does not have enough arguments. |
ERR 105 |
|
# ARGMTS Command does not have enough arguments. |
ERR 106 |
|
ABORTED! Current operation is aborted; GPIB parser reset. |
ERR 107 |
|
HELLO ?? No GPIB listener is present. |
ERR 108 |
|
TIME OUT Analyzer timed out when acting as controller. |
ERR 109 |
|
CtrlFail Analyzer unable to take control of the bus. |
ERR 110 |
|
NOT CTRL Analyzer is not system controller. |
ERR 111 |
|
# ARGMTS Command does not have enough arguments. |
ERR 112 |
|
??CMD?? Unrecognized command. |
ERR 113 |
|
FREQ NO! Command cannot have frequency units. |
ERR 114 |
|
TIME NO! Command cannot have time units. |
ERR 115 |
|
AMPL NO! Command cannot have amplitude units. |
ERR 116 |
|
?UNITS?? Unrecognizable units. |
ERR 117 |
|
NOP NUM Command cannot have numeric units. |
ERR 118 |
|
NOP EP Enable parameter cannot be used. |
ERR 119 |
|
NOP UPDN UP/DN are not valid arguments for this command. |
ERR 120 |
|
NOP ONOF ON/OFF are not valid arguments for this command. |
ERR 121 |
|
NOP ARG AUTO/MAN are not valid arguments for this command. |
ERR 122 |
|
NOP TRC Trace registers are not valid for this command. |
ERR 123 |
|
NOP ABLK A-block format not valid here. |
ERR 124 |
|
NOP IBLK I-block format not valid here. |
ERR 125 |
|
NOP STRNG Strings are not valid for this command. |
ERR 126 |
|
NO ? This command cannot be queried. |
ERR 127 |
|
BAD DTMD Not a valid peak detector mode. |
ERR 128 |
|
PK WHAT? Not a valid peak search parameter. |
ERR 129 |
|
PRE TERM Premature A-block termination. |
ERR 130 |
|
BAD TDF Arguments are only for TDF command. |
ERR 131 |
|
?? AM/FM AM/FM are not valid arguments for this command. |
ERR 132 |
|
!FAV/RMP FAV/RAMP are not valid arguments for this command. |
ERR 133 |
|
!INT/EXT INT/EXT are not valid arguments for this command. |
ERR 134 |
|
??? ZERO ZERO is not a valid argument for this command. |
ERR 135 |
|
??? CURR CURR is not a valid argument for this command. |
ERR 136 |
|
??? FULL FULL is not a valid argument for this command. |
ERR 137 |
|
??? LAST LAST is not a valid argument for this command. |
ERR 138 |
|
!GRT/DSP GRT/DSP are not valid arguments for this command. |
ERR 139 |
|
PLOTONLY Argument can only be used with PLOT command. |
ERR 140 |
|
?? PWRON PWRON is not a valid argument for this command. |
ERR 141 |
|
BAD ARG Argument can only be used with FDIAG command. |
ERR 142 |
|
BAD ARG Query expected for FDIAG command. |
ERR 143 |
|
NO PRESL No preselector hardware to use command with. |
ERR 144 |
|
COUPL?? Invalid COUPLING argument, expected AC or DC. |
ERR 200 |
|
SYSTEM ADC Driver/ADC hardware/firmware interaction; check for other errors. |
ERR 201 |
|
SYSTEM ADC Controller/ADC hardware/firmware interaction; check for other errors. |
ERR 202 |
|
FADC CAL Linear offset search failed. This error applies applies only to EC-series analyzers and E-series analyzers with fast ADC (Option 007) |
ERR 203 |
|
FADC CAL Log offset search failed. This error applies applies only to EC-series analyzers and E-series analyzers with fast ADC (Option 007) |
ERR 204 |
|
FADC CAL Log 2´ expand offset search failed. This error applies applies only to EC-series analyzers and E-series analyzers with fast ADC (Option 007) |
ERR 205 |
|
FADC CAL Linear slope calibration is out of range. This error applies applies only to EC-series analyzers and E-series analyzers with fast ADC (Option 007) |
ERR 206 |
|
FADC CAL Log slope calibration is out of range. This error applies applies only to EC-series analyzers and E-series analyzers with fast ADC (Option 007) |
ERR 207 |
|
FADC CAL Log 2´ expand slope calibration is out of range. This error applies applies only to EC-series analyzers and E-series analyzers with fast ADC (Option 007) |
ERR 250 |
|
OUTOF RG ADC input is outside of ADC range. |
ERR 251 |
|
NO IRQ Microprocessor not receiving interrupt from ADC. |
ERR 300 |
|
YTO UNLK YTO (1ST LO) phase-locked loop (PLL) is unlocked. |
ERR 301 |
|
YTO UNLK YTO PLL is unlocked. |
ERR 302 |
|
OFF UNLK Offset roller oscillator PLL is unlocked. |
ERR 303 |
|
XFR UNLK Transfer roller oscillator PLL is unlocked. |
ERR 304 |
|
ROL UNLK Main roller oscillator PLL is unlocked. |
ERR 305 |
|
FREQ ACC Coarse adjust DAC cannot bring MAINSENSE close to zero. |
ERR 306 |
|
FREQ ACC Fine adjust DAC cannot bring MAINSENSE close to zero. |
ERR 307 |
|
FREQ ACC Transfer oscillator pretune DAC out of range. |
ERR 308 |
|
FREQ ACC Offset oscillator pretune DAC not within prescribed limits at low frequency. |
ERR 309 |
|
FREQ ACC Offset oscillator pretune DAC not within prescribed limits at high frequency. |
ERR 310 |
|
FREQ ACC Main roller pretune DAC set to 255 before MAINSENSE changes to negative polarity. |
ERR 311 |
|
FREQ ACC Main roller pretune DAC set to 255 before MAINSENSE changes to negative polarity. |
ERR 312 |
|
FREQ ACC Fine adjust DAC cannot bring MAINSENSE close to zero. |
ERR 313 |
|
FREQ ACC The combination of sampler oscillator and roller oscillator frequencies do not correspond to the required YTO start frequency. |
ERR 314 |
|
FREQ ACC Span calibration problems. An unlocked main roller loop or lack of a sweep ramp. |
ERR 315 |
|
FREQ ACC Span calibration problems. Roller span attenuator DAC is out of range. |
ERR 316 |
|
FREQ ACC Sensitivity of the main roller DAC is zero. |
ERR 317 |
|
FREQ ACC Main coil coarse DAC at limits. |
ERR 318 |
|
FREQ ACC Main coil fine DAC at limits. |
ERR 319 |
|
WARN COA The YTO coarse tune DAC is near its limits. |
ERR 320 |
|
WARN FIN The YTO fine tune DAC is near its limits. |
ERR 321 |
|
FREQ ACC Main roller tuning sensitivity is not greater than zero. |
ERR 322 |
|
FREQ ACC Main roller pretune DAC value set greater than 255. |
ERR 324 |
|
FREQ ACC Coarse adjust DAC cannot bring MAINSENSE close to zero. |
ERR 325 |
|
FREQ ACC Fine adjust DAC cannot bring MAINSENSE close to zero. |
ERR 326 |
|
FREQ ACC Fine adjust DAC near the end of range. |
ERR 327 |
|
OFF UNLK Offset roller oscillator PLL is unlocked. |
ERR 328 |
|
FREQ ACC Roller fine adjust DAC sensitivity less than or equal to zero. |
ERR 329 |
|
FREQ ACC Roller coarse adjust DAC sensitivity less than or equal to zero. |
ERR 331 |
|
FREQ ACC Invalid YTO frequency. |
ERR 333 |
|
600 UNLK 600 MHz reference oscillator PLL is unlocked. |
ERR 334 |
|
LO AMPL 1st LO distribution amplifier is unleveled. The front panel LO OUTPUT must be terminated in a 50W load. |
ERR 335 |
|
SMP UNLK Sampling oscillator PLL is unlocked. |
ERR 336 |
|
10 MHz Ref Cal oscillator failed to lock when going to internal 10 MHz reference. |
ERR 337 |
|
FN UNLK Fractional N circuitry is unable to lock. |
ERR 338 |
|
LOMA AGC The LOMA (local oscillator multiplier amplifier) main loop AGC is unleveled. |
ERR 339 |
|
SBTX AGC The LOMA SBTX (switched barium-tuned mixer) loop AGC is unleveled. |
ERR 340 |
|
SAMP AGC The LOMA (local oscillator multiplier amplifier) sampler loop AGC is unleveled. |
ERR 351 |
|
SETL FLD The YTO error voltage is not settling. |
ERR 352 |
|
TWID FLD Unable to bring error voltage DACs to quiescent point. |
ERR 353 |
|
SRCH FLD No acceptable YTO DAC value found. |
ERR 354 |
|
LK ITERS Cannot lock. Lock iteration routine terminated. |
ERR 355 |
|
SMP CAL Sampler unlock condition during calibration routine. This error remains until a successful recalibration is performed. |
ERR 356 |
|
SPAC CAL Sweep data problem finding "bucket 1" of the span accuracy calibration sweep. |
ERR 357 |
|
SPAC CAL Cannot find the "x" intersection for "bucket 1" of the span accuracy calibration sweep. |
ERR 358 |
|
SPAC CAL Cannot find "bucket 2" of the span accuracy calibration sweep. |
ERR 359 |
|
SPAC CAL Cannot find the "x" intersection for "bucket 2" of the span accuracy calibration sweep. |
ERR 360 |
|
SPAC CAL The start bucket correction is out of range. |
ERR 361 |
|
SPAC CAL The percent of span correction is out of range. |
ERR 400 |
|
AMPL<300 Unable to adjust less than 300 Hz RES BW amplitude. |
ERR 401 |
|
AMPL 300 Unable to adjust amplitude of 300 Hz RES BW. |
ERR 402 |
|
AMPL 1K Unable to adjust amplitude of 1 kHz RES BW. |
ERR 403 |
|
AMPL 3K Unable to adjust amplitude of 3 kHz RES BW. |
ERR 404 |
|
AMPL 10K Unable to adjust amplitude of 10 kHz RES BW. |
ERR 405 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 1. |
ERR 406 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 2. |
ERR 407 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 3. |
ERR 408 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 4. |
ERR 409 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 1. |
ERR 410 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 2. |
ERR 411 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 3. |
ERR 412 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 4. |
ERR 413 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 1. |
ERR 414 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 2. |
ERR 415 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 3. |
ERR 416 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 4. |
ERR 417 |
|
RBW 3K Unable to adjust 3 kHz RES BW pole 1. |
ERR 418 |
|
RBW 3K Unable to adjust 3 kHz RES BW pole 2. |
ERR 419 |
|
RBW 3K Unable to adjust 3 kHz RES BW pole 3. |
ERR 420 |
|
RBW 3K Unable to adjust 3 kHz RES BW pole 4. |
ERR 421 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 1. |
ERR 422 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 2. |
ERR 423 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 3. |
ERR 424 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 4. |
ERR 425 |
|
RBW 3K Unable to adjust 3 kHz RES BW pole 1. |
ERR 426 |
|
RBW 3K Unable to adjust 3 kHz RES BW pole 2. |
ERR 427 |
|
RBW 3K Unable to adjust 3 kHz RES BW pole 3. |
ERR 428 |
|
RBW 3K Unable to adjust 3 kHz RES BW pole 4. |
ERR 429 |
|
RBW <300 Unable to adjust less than 300 Hz RES BW. |
ERR 430 |
|
RBW 300 Unable to adjust 300 Hz RES BW. |
ERR 431 |
|
RBW 1K Unable to adjust 1 kHz RES BW. |
ERR 432 |
|
RBW 3K Unable to adjust 3 kHz RES BW. |
ERR 433 |
|
RBW 10K Unable to adjust 10 kHz RES BW. |
ERR 434 |
|
RBW 300 Unable to adjust 300 Hz RES BW pole 1. |
ERR 435 |
|
RBW 300 Unable to adjust 300 Hz RES BW pole 2. |
ERR 436 |
|
RBW 300 Unable to adjust 300 Hz RES BW pole 3. |
ERR 437 |
|
RBW 300 Unable to adjust 300 Hz RES BW pole 4. |
ERR 438 |
|
RBW 1K Unable to adjust 1 kHz RES BW pole 1. |
ERR 439 |
|
RBW 1K Unable to adjust 1 kHz RES BW pole 2. |
ERR 440 |
|
RBW 1K Unable to adjust 1 kHz RES BW pole 3. |
ERR 441 |
|
RBW 1K Unable to adjust 1 kHz RES BW pole 4. |
ERR 442 |
|
RBW 3K Unable to adjust 3 kHz RES BW pole 1. |
ERR 443 |
|
RBW 3K Unable to adjust 3 kHz RES BW pole 2. |
ERR 444 |
|
RBW 3K Unable to adjust 3 kHz RES BW pole 3. |
ERR 445 |
|
RBW 3K Unable to adjust 3 kHz RES BW pole 4. |
ERR 446 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 1. |
ERR 447 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 2. |
ERR 448 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 3. |
ERR 449 |
|
RBW 10K Unable to adjust 10 kHz RES BW pole 4. |
ERR 450 |
|
IF SYSTM IF hardware failure. Check other error messages. |
ERR 451 |
|
IF SYSTM IF hardware failure. Check other error messages. |
ERR 452 |
|
IF SYSTM IF hardware failure. Check other error messages. |
ERR 454 |
|
AMPL Unable to adjust step gain amplifiers. |
ERR 455 |
|
AMPL Unable to adjust step gain amplifiers. |
ERR 456 |
|
AMPL Unable to adjust step gain amplifiers. |
ERR 457 |
|
AMPL Unable to adjust step gain amplifiers. |
ERR 458 |
|
AMPL Unable to adjust step gain amplifiers. |
ERR 459 |
|
AMPL Unable to adjust step gain amplifiers. |
ERR 460 |
|
AMPL Unable to adjust step gain amplifiers. |
ERR 461 |
|
AMPL Unable to adjust step gain amplifiers. |
ERR 462 |
|
AMPL Unable to adjust step gain amplifiers. |
ERR 463 |
|
AMPL Unable to adjust step gain amplifiers. |
ERR 464 |
|
AMPL Unable to adjust step gain amplifiers. |
ERR 465 |
|
AMPL Unable to adjust step gain amplifiers. |
ERR 466 |
|
LIN AMPL Unable to adjust linear amplitude scale. |
ERR 467 |
|
LOG AMPL Unable to adjust log amplitude scale. |
ERR 468 |
|
LOG AMPL Unable to adjust log amplitude scale. |
ERR 469 |
|
LOG AMPL Unable to adjust log amplitude scale. |
ERR 470 |
|
LOG AMPL Unable to adjust log amplitude scale. |
ERR 471 |
|
RBW 30K Unable to adjust 30 kHz RES BW. |
ERR 472 |
|
RBW 100K Unable to adjust 100 kHz RES BW. |
ERR 473 |
|
RBW 300K Unable to adjust 300 kHz RES BW. |
ERR 474 |
|
RBW 1M Unable to adjust 1 MHz RES BW. |
ERR 475 |
|
RBW 30K Unable to adjust 30 kHz RES BW. |
ERR 476 |
|
RBW 100K Unable to adjust 100 kHz RES BW. |
ERR 477 |
|
RBW 300K Unable to adjust 300 kHz RES BW. |
ERR 478 |
|
RBW 1M Unable to adjust 1 MHz RES BW. |
ERR 483 |
|
RBW 10K Unable to adjust 10 kHz RES BW. |
ERR 484 |
|
RBW 3K Unable to adjust 3 kHz RES BW. |
ERR 485 |
|
RBW 1K Unable to adjust 1 kHz RES BW. |
ERR 486 |
|
RBW 300 Unable to adjust 300 Hz RES BW. |
ERR 487 |
|
RBW 100 Unable to adjust 100 Hz RES BW. |
ERR 488 |
|
RBW 100 Unable to adjust 100 Hz RES BW. |
ERR 489 |
|
RBW 100 Unable to adjust 100 Hz RES BW. |
ERR 490 |
|
RBW 100 Unable to adjust 100 Hz RES BW. |
ERR 491 |
|
RBW <300 Unable to adjust less than 300 Hz RES BW. |
ERR 492 |
|
RBW 300 Unable to adjust 300 Hz RES BW. |
ERR 493 |
|
RBW 1K Unable to adjust 1 kHz RES BW. |
ERR 494 |
|
RBW 3K Unable to adjust 3 kHz RES BW. |
ERR 495 |
|
RBW 10K Unable to adjust 10 kHz RES BW. |
ERR 496 |
|
RBW 100 Unable to adjust 100 Hz RES BW. |
ERR 497 |
|
RBW 100 Unable to adjust 100 Hz RES BW. |
ERR 498 |
|
RBW 100 Unable to adjust 100 Hz RES BW. |
ERR 499 |
|
CAL UNLK Cal oscillator is unlocked. |
ERR 500 |
|
AMPL 30K Unable to adjust amplitude of 30 kHz RES BW. |
ERR 501 |
|
AMPL .1M Unable to adjust amplitude of 100 kHz RES BW. |
ERR 502 |
|
AMPL .3M Unable to adjust amplitude of 300 kHz RES BW. |
ERR 503 |
|
AMPL 1M Unable to adjust amplitude of 1 MHz RES BW. |
ERR 504 |
|
AMPL 30K Unable to adjust amplitude of 30 kHz RES BW. |
ERR 505 |
|
AMPL .1M Unable to adjust amplitude of 100 kHz RES BW. |
ERR 506 |
|
AMPL .3M Unable to adjust amplitude of 300 kHz RES BW. |
ERR 507 |
|
AMPL 1M Unable to adjust amplitude of 1 MHz RES BW. |
ERR 508 |
|
AMPL 30K Insufficient gain during LC BW Cal of 30 kHz RES BW. |
ERR 509 |
|
AMPL .1M Insufficient gain during LC BW Cal of 100 kHz RES BW. |
ERR 510 |
|
AMPL .3M Insufficient gain during LC BW Cal of 300 kHz RES BW. |
ERR 511 |
|
AMPL 1M Insufficient gain during LC BW Cal of 1 MHz RES BW. |
ERR 512 |
|
RBW <300 Insufficient gain during crystal BW Cal of less than 300 Hz RES BW. |
ERR 513 |
|
RBW 300 Insufficient gain during crystal BW Cal of 300 Hz RES BW. |
ERR 514 |
|
RBW 1K Insufficient gain during crystal BW Cal of 1 kHz RES BW. |
ERR 515 |
|
RBW 3K Insufficient gain during crystal BW Cal of 3 kHz RES BW. |
ERR 516 |
|
RBW 10K Insufficient gain during crystal BW Cal of 10 kHz RES BW. |
ERR 517 |
|
RBW 100 Unable to adjust 100 Hz RES BW. |
ERR 518 |
|
RBW 300 Unable to adjust 300 Hz RES BW. |
ERR 519 |
|
RBW 1K Unable to adjust 1 kHz RES BW. |
ERR 520 |
|
RBW 3K Unable to adjust 3 kHz RES BW. |
ERR 521 |
|
RBW 10K Unable to adjust 10 kHz RES BW. |
ERR 522 |
|
RBW 10K Unable to adjust 10 kHz RES BW symmetry of pole 1. |
ERR 523 |
|
RBW 10K Unable to adjust 10 kHz RES BW symmetry of pole 2. |
ERR 524 |
|
RBW 10K Unable to adjust 10 kHz RES BW symmetry of pole 3. |
ERR 525 |
|
RBW 10K Unable to adjust 10 kHz RES BW symmetry of pole 4. |
ERR 526 |
|
RBW <300 Unable to adjust less than 300 Hz RES BWs. Timeout during data sampling. |
ERR 527 |
|
RBW <300 Step gain correction failed for less than 300 Hz RES BW. |
ERR 528 |
|
RBW <300 Unable to adjust less than 300 Hz RES BWs. DC level at ADC cannot be calibrated. |
ERR 529 |
|
RBW <300 Unable to adjust less than 300 Hz RES BWs. Demod data for calibration is bad. |
ERR 530 |
|
RBW <300 Unable to adjust less than 300 Hz RES BWs. Narrow BW VCXO Calibration failed. |
ERR 531 |
|
RBW <300 Flatness correction data not acceptable for less than 300 Hz RES BWs. |
ERR 532 |
|
RBW <300 Absolute gain data for less than 300 Hz RES BWs not acceptable. |
ERR 533 |
|
RBW <300 Unable to adjust less than 300 Hz RES BWs. Timeout during data sampling narrow BW chunk. |
ERR 534 |
|
RBW <300 Unable to adjust less than 300 Hz RES BWs. Frequency count of CAL OSC using IF downconverter failed. |
ERR 535 |
|
RBW <300 Unable to adjust less than 300 Hz RES BWs. Inadequate FM demod range for 500Hz IF filter measurement. |
ERR 536 |
|
RBW <300 Unable to adjust less than 300 Hz RES BWs. Unable to autorange the chirp signal. |
ERR 537 |
|
RBW <300 Unable to adjust less than 300 Hz RES BWs. CW CAL OSC not autoranging. |
ERR 538 |
|
RBW <300 Unable to adjust less than 300 Hz RES BWs. Noisy shape of 500 Hz IF filter. |
ERR 539 |
|
RBW <300 Unable to adjust less than 300 Hz RES BWs. CW CAL OSC not autoranging. |
ERR 540 |
|
RBW <300 Unable to adjust less than 300 Hz RES BWs. No SIGLOCK on CW CAL OSC during pretune. |
ERR 550 |
|
IDCALOSC CAL Oscillator ID. Indicates incompatible hardware. Cal oscillator not expected. |
ERR 551 |
|
ID LOGBD LOG Board ID. Indicates incompatible hardware. Log board not expected. |
ERR 552-562 |
|
LOG AMPL Unable to adjust amplitude of log scale. |
ERR 563 |
|
LOG AMPL Unable to adjust amplitude in log scale. Third Step Gain range problem. |
ERR 564-570 |
|
LOG AMPL Unable to adjust amplitude in log scale. |
ERR 571 |
|
AMPL Unable to adjust step gain amplifiers. |
ERR 572 |
|
AMPL 1M Unable to adjust amplitude of 1 MHz RES BW. |
ERR 573-577 |
|
LOG AMPL Unable to adjust amplitude in log scale. |
ERR 578 |
|
LOG AMPL Limiter calibration error from DC logger calibration. |
ERR 579 |
|
LOG AMPL Attenuator CAL level error from DC logger calibration. |
ERR 580 |
|
LOG AMPL Calibration level error from DC logger fidelity calibration. |
ERR 581 |
|
AMPL Unable to adjust 100 kHz resolution bandwidth and resolution bandwidths less than or equal to 10 kHz. |
ERR 582 |
|
AMPL Unable to adjust 100 kHz resolution bandwidth and resolution bandwidths less than or equal to 10 kHz. Bad CALOSC calibration of sweep rate. |
ERR 583 |
|
RBW 30K Unable to adjust 30 kHz RES BW. |
ERR 584 |
|
RBW 100K Unable to adjust 100 kHz RES BW. |
ERR 585 |
|
RBW 300K Unable to adjust 300 kHz RES BW. |
ERR 586 |
|
RBW 1M Unable to adjust 1 MHz RES BW. |
ERR 587 |
|
RBW 30K Unable to adjust 30 kHz RES BW. |
ERR 588 |
|
RBW 100K Unable to adjust 100 kHz RES BW. |
ERR 589 |
|
RBW 300K Unable to adjust 300 kHz RES BW. |
ERR 590 |
|
RBW 1M Unable to adjust 1 MHz RES BW. |
ERR 591 |
|
LOG AMPL Unable to adjust amplitude in log scale. |
ERR 592 |
|
LOG AMPL Unable to adjust amplitude in log scale. |
ERR 593 |
|
LOG TUNE LIMIT Cal tune error from DC logger calibration. |
ERR 594 |
|
LOG OFST Attenuator calibration offset error from DC logger calibration. |
ERR 595 |
|
LOG ATTN Attenuator calibration absolute error from DC logger calibration. |
ERR 596 |
|
LOG FID Fidelity error from DC logger calibration. |
ERR 597 |
|
LOG OFST Fidelity offset error from DC logger calibration. |
ERR 598 |
|
LOG OFST Fidelity offset unstable from DC logger calibration. |
ERR 599 |
|
LOG GAIN Fidelity gain error from DC logger calibration. |
ERR 600 |
|
SYSTEM Hardware/firmware interaction; check other errors. |
ERR 601 |
|
SYSTEM Hardware/firmware interaction; check other errors. |
ERR 650 |
|
OUTOF RG ADC input is outside of the ADC range. |
ERR 651 |
|
NO IRQ Microprocessor is not receiving interrupt from ADC. |
ERR 700 |
|
EEROM Checksum error of EEROM A2U501. |
ERR 701 |
|
AMPL CAL Checksum error of frequency response correction data. |
ERR 702 |
|
ELAP TIM Checksum error of elapsed time data. |
ERR 703 |
|
AMPL CAL Checksum error of frequency response correction data. |
ERR 704 |
|
PRESELCT Checksum error of customer preselector peak data; external preselector data recalled in internal mode; or internal preselector data recalled in external mode. |
ERR 705 |
|
ROM U306 Checksum error of program ROM A2U306. |
ERR 706 |
|
ROM U307 Checksum error of program ROM A2U307. |
ERR 707 |
|
ROM U308 Checksum error of program ROM A2U308. |
ERR 708 |
|
ROM U309 Checksum error of program ROM A2U309. |
ERR 709 |
|
ROM U310 Checksum error of program ROM A2U310. |
ERR 710 |
|
ROM U311 Checksum error of program ROM A2U311. |
ERR 711 |
|
RAM U303 Checksum error of system RAM A2U303. |
ERR 712 |
|
RAM U302 Checksum error of system RAM A2U302. |
ERR 713 |
|
RAM U301 Checksum error of system RAM A2U301. |
ERR 714 |
|
RAM U300 Checksum error of system RAM A2U300. |
ERR 715 |
|
RAM U305 Checksum error of system RAM A2U305. |
ERR 716 |
|
RAM U304 Checksum error of system RAM A2U304. |
ERR 717 |
|
BAD uP!! Microprocessor is not fully operational. |
ERR 718 |
|
BATTERY? Nonvolatile RAM is not working; check battery. |
ERR 719 |
|
MODEL #? Cannot identify the model number. |
ERR 720 |
|
OPTION #? Cannot identify an option number. |
ERR 721 |
|
AMPC RAM Checksum error of the ampcor function correction data. |
ERR 750 |
|
SYSTEM Hardware/firmware interaction, zero divide; check other errors. |
ERR 751 |
|
SYSTEM Hardware/firmware interaction, floating overflow; check other errors. |
ERR 752 |
|
SYSTEM Hardware/firmware interaction, floating underflow; check other errors. |
ERR 753 |
|
SYSTEM Hardware/firmware interaction, LOG error; check other errors. |
ERR 754 |
|
SYSTEM Hardware/firmware interaction, Integer overflow; check other errors. |
ERR 755 |
|
SYSTEM Hardware/firmware interaction, squareroot error; check other errors. |
ERR 756 |
|
SYSTEM Hardware/firmware interaction, Triple overflow; check other errors. |
ERR 757 |
|
SYSTEM Hardware/firmware interaction, BCDL overflow; check other errors. |
ERR 758 |
|
SYSTEM Unknown system error. |
ERR 759 |
|
SYSTEM Hardware/firmware interaction. Code invoked for wrong instrument. |
ERR 760 |
|
NO FADC The power-on read of the fast analog to digital converter hardware is failing. This error applies applies only to EC-series analyzers and E-series analyzers with fast ADC (Option 007) |
ERR 900 |
|
TG UNLVL Tracking generator output is unleveled. |
ERR 901 |
|
TGFrqLmt Tracking generator output unleveled because START FREQ is set below tracking generator frequency limit (300 kHz). |
ERR 902 |
|
BAD NORM A normalization error will occur if the current spectrum analyzer state is not the same as the state stored by the last execution of the STORETHRU or STORESHORT command. |
ERR 903 |
|
A > DLMT Unnormalized trace A is off screen with trace math or normalization on. |
ERR 904 |
|
B > DLMT Calibration trace (trace B) is off screen with trace math or normalization on. |
ERR 905 |
|
EXT REF Unable to lock cal oscillator when set to external reference. Check that external reference is within tolerance. |
ERR 906 |
|
OVENCOLD The oven controlled crystal oscillator (OCXO) oven is cold. |
ERR 907 |
|
DO IFCAL Unit is still performing IF calibration, or is in need of IF calibration which were not done due to an OVENCOLD condition. |
ERR 908 |
|
BW>>SPCG Channel bandwidth is too wide, compared to the channel spacing, for a meaningful computation. |
ERR 909 |
|
SPAN<ACP The frequency span is too small to obtain a valid ACP measurement. |
ERR 910 |
|
SPAN>ACP The frequency span is too wide, compared to the channel bandwidth, to obtain a valid ACP measurement. |
ERR 911 |
|
ACP STATE The adjacent channel power measurement has been compromised (invalid measurement parameters). |
ERR 919 |
|
SP <CHBW The channel spacing is too narrow, compared to the channel bandwidth, to obtain a valid channel power bandwidth measurement. |
ERR 920 |
|
RBW>CHBW The resolution bandwidth is too wide, compared to the channel bandwidth, to obtain a valid channel power bandwidth measurement. The resolution bandwidth should be much less than the channel bandwidth (<0.1´channel BW). |
ERR 921 |
|
AMPCOR Measurement data which would normally be displayed above the top graticule, and therefore has unspecified accuracy, has been corrected by ampcor the amplitude correction function (ampcor) to appear between the top and bottom graticule. |
ERR 922 |
|
¯AMPCOR¯ Measurement data which would normally be displayed below the bottom graticule, and therefore has unspecified accuracy, has been corrected by ampcor the amplitude correction function (ampcor) to appear between the top and bottom graticule. |