68 |
ADDITIONAL STANDARDS NEEDED |
|
Error correction for the selected calibration class cannot be computed until you have measured all the necessary standards. |
31 |
ADDRESSED TO TALK WITH NOTHING TO SAY |
|
You have sent a read command to the analyzer (such as ENTER 716) without first requesting data with an appropriate output command (such as OUTPDATA). The analyzer has no data in the output queue to satisfy the request. |
200 |
ALL REGISTERS HAVE BEEN USED |
|
You have used all of the available registers; you can store no more instrument states even though you may still have sufficient memory. There are 31 registers available, plus the preset instrument state. |
212 |
ANALOG BUS DISABLED IN 6 KHZ IF BW |
|
When you press Avg, IF BW [6000], the analog bus is disabled and not available for use in troubleshooting. For a description of the analog bus, refer to the service guide. |
60 |
ANALOG INPUT OVERLOAD |
|
The power level of the analog input is too high. Reduce the power level of the analog input source. |
37 |
ANOTHER SYSTEM CONTROLLER ON GPIB |
|
You must remove the active controller from the bus or the controller must relinquish the bus before the analyzer can assume the system controller mode. |
206 |
ARGUMENT OUT OF RANGE |
|
The argument for a programming command is out of the specified range. Refer to the programmer’s guide for a list of programming commands and argument ranges. |
193 |
ASCII: MISSING 'BEGIN' STATEMENT |
|
The CITIfile you just downloaded over the GPIB or via disk was not properly organized. The analyzer is unable to read the “BEGIN” statement. |
194 |
ASCII: MISSING 'CITIFILE' STATEMENT |
|
The CITIfile you just downloaded over the GPIB or via disk was not properly organized. The analyzer is unable to read the “CITIFILE” statement. |
195 |
ASCII: MISSING 'DATA' STATEMENT |
|
The CITIfile you just downloaded over the GPIB or via disk was not properly organized. The analyzer is unable to read the “DATA” statement. |
196 |
ASCII: MISSING 'VAR' STATEMENT |
|
The CITIfile you just downloaded over the GPIB or via disk was not properly organized. The analyzer is unable to read the “VAR” statement. |
13 |
AVERAGING INVALID ON NON-RATIO MEASURE |
|
You cannot use sweep-to-sweep averaging in single-input measurements. Sweep-to-sweep averaging is valid only for ratioed measurements (A/R, B/R, A/B, and S-parameters). You can use other noise reduction techniques, such as narrower IF bandwidth, for single input measurements. |
181 |
BAD FREQ FOR HARMONIC |
|
You recalled a calibration that resulted in start and stop frequencies that are beyond the allowable limits. |
183 |
BATTERY FAILED. STATE MEMORY CLEARED |
|
The battery protection of the non-volatile CMOS memory has failed. The CMOS memory has been cleared. Refer to Chapter 8, “Preset State and Memory Allocation” for more information about the CMOS memory. |
184 |
BATTERY LOW! STORE SAVE REGS TO DISK |
|
The battery protection of the non-volatile CMOS memory is in danger of failing. If this occurs, all of the instrument state registers stored in CMOS memory will be lost. Save these states to a disk. Refer to Chapter 8, “Preset State and Memory Allocation” for more information about the CMOS memory. |
34 |
BLOCK INPUT ERROR |
|
The analyzer did not receive a complete data transmission. This is usually caused by an interruption of the bus transaction. Clear by pressing the Local, key or aborting the I/O process at the controller. |
35 |
BLOCK INPUT LENGTH ERROR |
|
The length of the header received by the analyzer did not agree with the size of the internal array block. Refer to the programmer’s guide for instructions on using analyzer input commands. |
74 |
CALIBRATION ABORTED |
|
You have changed the active channel during a calibration, so the calibration in progress was terminated. Make sure the appropriate channel is active and restart the calibration. |
63 |
CALIBRATION REQUIRED |
|
A calibration set could not be found that matched the current stimulus state or measurement parameter. You will have to perform a new calibration. |
185 |
CANNOT FORMAT DOS DISKS ON THIS DRIVE |
|
You have attempted to initialize a floppy disk to DOS format on an external disk drive that does not support writing to all 80 tracks of the double density and high density disks. The older single-sided disks had only 66 tracks and some disk drives were limited to accessing that number of tracks. To format the disk, either choose another external disk drive or use the analyzer's internal disk drive. |
199 |
CANNOT MODIFY FACTORY PRESET |
|
You have attempted to rename, delete, or otherwise alter the factory preset state. The factory preset state is permanently stored in ROM and cannot be altered. If your intent was to create a user preset state, you must create a new instrument state, save it, and then rename it to “UPRESET”. Refer to Chapter 8, “Preset State and Memory Allocation” for more detailed instructions. |
203 |
CANNOT READ/WRITE HFS FILE SYSTEM |
|
The disk is being accessed by the analyzer and is found to contain an HFS (hierarchical file system) or files nested within subdirectories. The analyzer does not support HFS. Replace the disk medium with a LIF or DOS formatted disk that does not contain files nested within subdirectories. |
127 |
CAN'T STORE/LOAD SEQUENCE, INSUFFICIENT MEMORY |
|
Your sequence transfer to or from a disk could not be completed due to insufficient memory. |
215 |
CAUTION: CORRECTION OFF: AUX CHANNEL(S) DISABLED |
|
This message is displayed when correction is forced off due to a stimulus change that is not compatible with the current calibration while an auxiliary channel is enabled. The auxiliary channels are restored when correction is turned on by pressing Cal, CORRECTION on OFF. |
218 |
CAUTION: FLOPPY DISK IS FULL |
|
This message is displayed if you try to save files to the floppy disk that does not have enough memory to perform the save task. Correct by inserting a new floppy disk in the front panel disk drive or by erasing files from the current floppy disk. |
159 |
CH1 (CH2, CH3, CH4) TARGET VALUE NOT FOUND |
|
Your target value for the marker search function does not exist on the current data trace. |
10 |
CONTINUOUS SWITCHING NOT ALLOWED |
|
Your current measurement requires different power ranges on channel 1 and channel 2. To protect the attenuator from undue mechanical wear, test sethold will be activated. The “tsH” (test set hold) indicator in the left margin of the display indicates that the inactive channel has been put in the sweep hold mode. This message is also displayed if a mechanical switch test set is in use and channels are measuring parameters that require the test set to switch continuously, for example S11 on Channel 1 and S22 on Channel 2 |
170 |
COPY: device not responding; copy aborted |
|
The printer or plotter is not accepting data. Verify the cable connections, GPIB addresses, and otherwise ensure that the copy device is ready. |
66 |
CORRECTION TURNED OFF |
|
Critical parameters in your current instrument state do not match the parameters for the calibration set, therefore correction has been turned off. The critical instrument state parameters are sweep type, start frequency, frequency span, and number of points. |
64 |
CURRENT PARAMETER NOT IN CAL SET |
|
Correction is not valid for your selected measurement parameter. Either change the measurement parameters or perform a new calibration. |
130 |
D2/D1 INVALID WITH SINGLE CHANNEL |
|
You can only make a D2/D1 measurement if both channels are on. |
152 |
D2/D1 INVALID: CH1 CH2 NUM PTS DIFFERENT |
|
You can only make a D2/D1 measurement if both channels have the same number of points. |
111 |
DEADLOCK |
|
A fatal firmware error occurred before instrument preset completed. Instrument needs service. |
17 |
DEMODULATION NOT VALID |
|
Demodulation was selected when the analyzer was not in CW time mode. Select demodulation only after putting the analyzer into CW time mode. |
119 |
DEVICE: not on, not connect, wrong addrs |
|
The device at the selected address cannot be accessed by the analyzer. Verify that the device is switched on, and check the GPIB connection between the analyzer and the device. Ensure that the device address recognized by the analyzer matches the GPIB address set on the device itself. |
188 |
DIRECTORY FULL |
|
There is no room left in the directory to add files. Either delete files or get a new disk. |
39 |
DISK HARDWARE PROBLEM |
|
The disk drive is not responding correctly. If using an external disk drive, refer to the disk drive operating manual. |
48 |
DISK IS WRITE PROTECTED |
|
The store operation cannot write to a write-protected disk. Slide the write-protect tab over the write-protect opening in order to write data on the disk. |
40 |
DISK MEDIUM NOT INITIALIZED |
|
You must initialize the disk before it can be used. |
190 |
DISK MESSAGE LENGTH ERROR |
|
The analyzer and the external disk drive aren't communicating properly. Check the GPIB connection and then try substituting another disk drive to isolate the problem instrument. |
38 |
DISK: not on, not connected, wrong addrs |
|
The disk cannot be accessed by the analyzer. Verify power to the disk drive, and check the GPIB connection between the analyzer and the disk drive. Ensure that the disk drive address recognized by the analyzer matches the GPIB address set on the disk drive itself. |
189 |
DISK READ/WRITE ERROR |
|
There may be a problem with your disk. Try a new floppy disk. If a new floppy disk does not eliminate the error, suspect hardware problems. |
49 |
DISK WEAR - REPLACE DISK SOON |
|
Cumulative use of the disk is approaching the maximum. Copy files as necessary using an external controller. If no controller is available, load instrument states from the old disk and store them to a newly initialized disk using the save/recall features of the analyzer. Discard the old disk. |
125 |
DUPLICATING TO THIS SEQUENCE NOT ALLOWED |
|
A sequence cannot be duplicated to itself. |
72 |
EXCEEDED 7 STANDARDS PER CLASS |
|
When modifying calibration kits, you can define a maximum of seven standards for any class. |
192 |
FILE NOT FOUND |
|
The requested file was not found on the current disk medium. |
197 |
FILE NOT FOUND OR WRONG TYPE |
|
During a recall operation, either the file was not found or the type of file was not an instrument state file. |
42 |
FIRST CHARACTER MUST BE A LETTER |
|
The first character of a disk file title or an internal save register title must be an alpha character. |
75 |
FORMAT NOT VALID FOR MEASUREMENT |
|
Conversion measurements (Z or Y reflection and transmission) are not valid with Smith chart and SWR formats. |
202 |
FUNCTION NOT AVAILABLE |
|
The function you requested over GPIB is not available on the current instrument. |
14 |
FUNCTION NOT VALID |
|
The function you requested is incompatible with the current instrument state. |
131 |
FUNCTION NOT VALID DURING MOD SEQUENCE |
|
You cannot perform sequencing operations while a sequence is being modified. |
201 |
FUNCTION NOT VALID FOR INTERNAL MEMORY |
|
The function you selected only works with disk files. |
163 |
FUNCTION ONLY VALID DURING MOD SEQUENCE |
|
You can only use the GOSUB SEQUENCE, capability when you are building a sequence. Attempting to use this softkey at any other time returns an error message and no action is taken. |
61 |
8703 SOURCE PARAMETERS CHANGED |
|
Some of the stimulus parameters of the instrument state have been changed, because you have turned correction on. A calibration set for the current measurement parameter was found and activated. The instrument state was updated to match the stimulus parameters of the calibration state. |
169 |
GPIB COPY IN PROGRESS, ABORT WITH LOCAL |
|
A GPIB copy was already in progress when you requested the GPIB for another function. To abort the first copy, press Local, otherwise the GPIB is unavailable until the first copy is completed. |
46 |
ILLEGAL UNIT OR VOLUME NUMBER |
|
The disk unit or volume number set in the analyzer is not valid. Refer to the disk drive operating manual. |
47 |
INITIALIZATION FAILED |
|
The disk initialization failed, probably because the disk is damaged. |
56 |
INSTRUMENT STATE MEMORY CLEARED |
|
All instrument state registers have been cleared from memory along with any saved calibration data, memory traces, and calibration kit definitions. Additionally, all user-settable selections (such as GPIB addresses) are set to their defaults. |
51 |
INSUFFICIENT MEMORY |
|
Your last front panel or GPIB request could not be implemented due to insufficient memory space. In some cases, this is a fatal error from which you can escape only by presetting the instrument. |
168 |
INSUFFICIENT MEMORY FOR PRINT/PLOT |
|
There is not enough memory available for the print or plot function. Increase the available memory by changing or eliminating a memory-intensive operation such as reducing the number of points in the sweep. |
154 |
INSUFFICIENT MEMORY, PWR MTR CAL OFF |
|
There is not enough memory space for the power meter calibration array. Increase the available memory by clearing one or more save/recall registers, or by reducing the number of points. |
2 |
INVALID KEY |
|
An invalid key was attempted; there may be a possible hardware failure. |
223 |
ISOL AVERAGES < SWP AVERAGES |
|
The isolation averages are less than the instrument sweep averages. Increase the isolation averages to be equal to or greater than instrument sweep averages. This error can only occur when instrument averaging is turned on. |
219 |
LASER TEMPERATURE LOOP OPEN |
|
The laser temperature is unstable. (This message may appear for the first few minutes when the analyzer is powered up.) Cycle the analyzer power and allow a warm-up time of two hours, then press any key. If the message persists, instrument needs service.. |
205 |
LIMIT TABLE EMPTY |
|
Limit lines cannot be turned on unless a limit table has been created. Refer to the “Making Measurements” chapter of the user’s guide for information on how to create a limit table. |
9 |
LIST TABLE EMPTY |
|
The frequency list is empty. To implement list frequency mode, add segments to the list table. |
150 |
LOG SWEEP REQUIRES 2 OCTAVE MINIMUM SPAN |
|
A logarithmic sweep is only valid if the stop frequency is greater than four times the start frequency. For frequency spans of less than two octaves, the sweep type automatically reverts to linear sweep. |
224 |
LW TESTSET MISSING |
|
The lightwave testset may be powered down separately, or the cable at the rear panel may be removed or loose. |
132 |
MEMORY FOR CURRENT SEQUENCE IS FULL |
|
All the memory in the sequence you are modifying is filled with instrument commands. |
220 |
MODULATOR TEMPERATURE LOOP OPEN |
|
The modulator temperature is unstable. (This message may appear for the first few minutes when the analyzer is powered up.) Cycle the analyzer power and allow a warm-up time of two hours, then press any key. If the message persists, instrument needs service.. |
71 |
MORE SLIDES NEEDED |
|
When you use a sliding load (in a user-defined calibration kit), you must set at least three slide positions to complete the calibration. |
69 |
NO CALIBRATION CURRENTLY IN PROGRESS |
|
The RESUME CAL SEQUENCE, softkey is not valid unless a calibration is already in progress. Start a new calibration. |
41 |
NO DISK MEDIUM IN DRIVE |
|
You have no disk in the current disk unit. Insert a disk, or check the disk unit number stored in the analyzer. |
114 |
NO FAIL FOUND |
|
The self-diagnose function of the instrument operates on an internal test failure. At this time, no failure has been detected. |
45 |
NO FILE(S) FOUND ON DISK |
|
No files of the type created by an analyzer store operation were found on the disk or the disk drive is empty. If you requested a specific file title, that file was not found on the disk. |
5 |
NO IF FOUND: CHECK R INPUT LEVEL |
|
The first IF signal was not detected during pretune. Check the front panel R channel jumper. If there is no visible problem with the jumper, refer to the service guide for troubleshooting. |
144 |
NO LIMIT LINES DISPLAYED |
|
You can turn limit lines on but they cannot be displayed on polar or Smith chart display formats. |
15 |
NO MARKER DELTA - SPAN NOT SET |
|
You must turn the delta marker mode on, with at least two markers displayed, in order to use the MARKER -> SPAN, softkey function. |
123 |
NO MEMORY AVAILABLE FOR INTERPOLATION |
|
You cannot perform interpolated error correction due to insufficient memory. |
126 |
NO MEMORY AVAILABLE FOR SEQUENCING |
|
You cannot modify the sequence due to insufficient memory. |
70 |
NO SPACE FOR NEW CAL. CLEAR REGISTERS |
|
You cannot store a calibration set due to insufficient memory. You can free more memory by clearing a saved instrument state from an internal register (which may also delete an associated calibration set, if all the instrument states using the calibration set have been deleted). You can store the saved instrument state and calibration set to a disk before clearing them. After deleting the instrument states, press Preset, to run the memory packer. |
198 |
NOT ALLOWED DURING POWER METER CAL |
|
When the analyzer is performing a power meter calibration, the GPIB bus is unavailable for other functions such as printing or plotting. |
44 |
NOT ENOUGH SPACE ON DISK FOR STORE |
|
The store operation will overflow the available disk space. Insert a new disk or purge files to create free disk space. |
54 |
NO VALID MEMORY TRACE |
|
If you are going to display or otherwise use a memory trace, you must first store a data trace to memory. |
55 |
NO VALID STATE IN REGISTER |
|
You have requested the analyzer, over GPIB (or by sequencing), to load an instrument state from an empty internal register. |
43 |
ONLY LETTERS AND NUMBERS ARE ALLOWED |
|
You can only use alpha-numeric characters (and underscores) in disk file titles or internal save register titles. Other symbols are not allowed, except for the “underscore” symbol. |
211 |
OVERLAP! LIST TYPE CHANGED TO STEPPED |
|
The list type changed to stepped because one or more frequency segments in the swept list table overlapped. Change the frequency ranges of the overlapping segments and switch back to swept list mode. |
58 |
OVERLOAD ON INPUT A, POWER REDUCED |
|
See OVERLOAD ON INPUT R, POWER REDUCED (error number 57). |
59 |
OVERLOAD ON INPUT B, POWER REDUCED |
|
See OVERLOAD ON INPUT R, POWER REDUCED (error number 57). |
57 |
OVERLOAD ON INPUT R, POWER REDUCED |
|
You have exceeded approximately +14 dBm at one of the test ports. The RF output power is automatically reduced to -85 dBm. The annotation P? appears in the left margin of the display to indicate that the power trip function has been activated. When this occurs, reset the power to a lower level, then toggle the softkey to switch on the power again. |
165 |
PARALLEL PORT NOT AVAILABLE FOR GPIO |
|
You have defined the parallel port as COPY for printing in the GPIB menu. To access the parallel port for general purpose I/O (GPIO), set the selection to PARALLEL [GPIO]. |
167 |
PARALLEL PORT NOT AVAILABLE FOR COPY |
|
You have defined the parallel port as general purpose I/O (GPIO) for sequencing. The definition was made under the Local, key menus. To access the parallel port for copy, set the selection to PARALLEL [COPY]. |
4 |
PHASE LOCK CAL FAILED |
|
An internal phase lock calibration routine is automatically executed at power-on, preset, and any time a loss of phase lock is detected. This message indicates that phase lock calibration was initiated and the first IF detected, but a problem prevented the calibration from completing successfully. |
7 |
PHASE LOCK FAILURE |
|
The first IF signal was detected at pretune, but phase lock could not be acquired. Check the signal level to the R channel input to make sure it is -35 dBm or higher. |
8 |
PHASE LOCK LOST |
|
Phase lock was acquired but then lost. |
27 |
PLOT ABORTED |
|
When you press the Local, key, the analyzer aborts the plot in progress. |
26 |
PLOTTER: not on, not connect, wrong addrs |
|
The plotter does not respond to control. Verify power to the plotter, and check the GPIB connection between the analyzer and the plotter. Ensure that the plotter address recognized by the analyzer matches the GPIB address set on the plotter itself. |
28 |
PLOTTER NOT READY-PINCH WHEELS UP |
|
The plotter pinch wheels clamp the paper in place. If you raise the pinch wheels, the plotter indicates a “not ready” status on the bus. |
6 |
POSSIBLE FALSE LOCK |
|
Phase lock has been achieved, but the source may be phase locked to the wrong harmonic of the synthesizer. |
116 |
POWER METER INVALID |
|
The power meter indicates an out-of-range condition. Check the test setup. |
118 |
POWER METER NOT SETTLED |
|
Sequential power meter readings are not consistent. Verify that the equipment is set up correctly. If so, preset the instrument and restart the operation. |
179 |
POWER OUT MAY BE UNLEVELED |
|
There is either a hardware failure in the source or you have attempted to set the power level too high. The analyzer allows the output power to be set higher or lower than the specified available power range. However, these output powers may be un-leveled or unavailable. Check to see if the power level you set is within specifications. |
21 |
POWER SUPPLY HOT! |
|
The temperature sensors on the A8 post-regulator assembly have detected an over-temperature condition. The power supplies regulated on the post-regulator have been shut down. |
22 |
POWER SUPPLY SHUT DOWN! |
|
One or more supplies on the A8 post-regulator assembly have been shut down due to an over-current, over-voltage, or under-voltage condition. |
25 |
PRINT ABORTED |
|
When you press the Local, key, the analyzer aborts output to the printer. |
178 |
print color not supported with EPSON |
|
You have defined the printer type as EPSON-P2. Color print is not supported with this printer. The print will abort. |
176 |
PRINTER: busy |
|
The parallel port printer is not accepting data. |
175 |
PRINTER: error |
|
The parallel port printer is malfunctioning. The analyzer cannot complete the copy function. |
173 |
PRINTER: not connected |
|
There is no printer connected to the parallel port. |
177 |
PRINTER: not handshaking |
|
The printer at the parallel port is not responding. |
172 |
PRINTER: not on line |
|
The printer at the parallel port is not set on line. |
24 |
PRINTER: not on, not connected, wrong addrs |
|
The printer does not respond to control. Verify power to the printer, and check the GPIB connection between the analyzer and the printer. Ensure that the printer address recognized by the analyzer matches the GPIB address set on the printer itself. |
171 |
PRINTER: paper error |
|
There is a paper-related problem with the parallel port printer such as a paper jam or out-of-paper condition. |
174 |
PRINTER: power off |
|
The power to the printer at the parallel port is off. |
166 |
PRINT/PLOT IN PROGRESS, ABORT WITH LOCAL |
|
If a print or plot is in progress and you attempt a second print or plot, this message is displayed and the second attempt is ignored. To abort a print or plot in progress, press Local. |
117 |
PWR MTR NOT ON/CONNECTED OR WRONG ADDRS |
|
The power meter cannot be accessed by the analyzer. Verify that the power meter address and model number set in the analyzer match the address and model number of the actual power meter. |
213 |
RANGE CAUSED POWER LVL CHANGE IN LIST |
|
The selected power range changed the power level of one or more segments in the swept list table. Change the segment power or change the power range. |
219 |
RECEIVER TEMPERATURE LOOP OPEN |
|
The receiver temperature is unstable. (This message may appear for the first few minutes when the analyzer is powered up.) Cycle the analyzer power and allow a warm-up time of two hours, then press any key. If the message persists, instrument needs service. |
30 |
REQUESTED DATA NOT CURRENTLY AVAILABLE |
|
The analyzer does not currently contain the data you have requested. For example, this condition occurs when you request error term arrays and no calibration is active. |
151 |
SAVE FAILED. INSUFFICIENT MEMORY |
|
You cannot store an instrument state in an internal register due to insufficient memory. Increase the available memory by clearing one or more save/recall registers and pressing Preset, or by storing files to a disk. |
124 |
SELECTED SEQUENCE IS EMPTY |
|
The sequence you attempted to run does not contain instrument commands. |
112 |
SELF TEST #n FAILED |
|
Internal test #n has failed. Several internal test routines are executed at instrument preset. The analyzer reports the first failure detected. |
157 |
SEQUENCE ABORTED |
|
The sequence running was stopped prematurely when you pressed the Local, key. |
153 |
SEQUENCE MAY HAVE CHANGED, CAN'T CONTINUE |
|
When you pause a sequence, you cannot continue it if you have modified it. You must start the sequence again. |
73 |
SLIDES ABORTED (MEMORY REALLOCATION) |
|
You cannot perform sliding load measurements due to insufficient memory. Increase the available memory by clearing one or more save/recall registers and pressing Preset, or by storing files to a disk and then deleting them from internal memory. |
187 |
SWEEP MODE CHANGED TO CW TIME SWEEP |
|
If you select external source auto or manual instrument mode and you do not also select CW mode, the analyzer is automatically switched to CW. |
11 |
SWEEP TIME INCREASED |
|
You have made instrument changes that cause the analyzer sweep time to be automatically increased. Some parameter changes that cause an increase in sweep time are narrower IF bandwidth, an increase in the number of points, and a change in sweep type. |
12 |
SWEEP TIME TOO FAST |
|
The fractional-N and digital IF circuits have lost synchronization. |
33 |
SYNTAX ERROR |
|
You have improperly formatted a GPIB command. Refer to the programmer’s guide for proper command syntax. |
36 |
SYST CTRL OR PASS CTRL IN LOCAL MENU |
|
The analyzer is in talker/listener mode. In this mode, the analyzer cannot control a peripheral device on the bus. Use the local menu to change to system controller or pass control mode. |
57 |
TEST PORT OVERLOAD, REDUCE POWER |
|
You have exceeded approximately +14 dBm at one of the test ports (or 0 dBm at the A or B sampler, Option 012 only). When this occurs, reduce the power to a lower level. |
133 |
THIS LIST FREQ INVALID |
|
You have set frequencies in the list that are outside of the allowable frequency range of the analyzer. Reduce the frequency range of the list. |
164 |
TOO MANY NESTED SEQUENCES. SEQ ABORTED |
|
You can only nest sequences to a maximum level of six. The sequence will abort if you nest more than six. |
50 |
TOO MANY SEGMENTS OR POINTS |
|
You can have a maximum of 30 segments or 1601 points in frequency list mode. In power meter calibrations, you can have a maximum of 12 segments for power sensor cal factors and power loss functions. |
32 |
WRITE ATTEMPTED WITHOUT SELECTING INPUT TYPE |
|
You have sent the data header “#A” to the analyzer with no preceding input command (such as INPUDATA). The instrument recognized the header but did not know what type of data to receive. Refer to the programmer’s guide for command syntax information. |
77 |
WRONG DISK FORMAT, INITIALIZE DISK |
|
You have attempted to store, load, or read file titles, but your disk format does not conform to the Logical Interchange Format (LIF) or DOS format. You must initialize the disk before reading or writing to it. |