68 |
ADDITIONAL STANDARDS NEEDED |
|
Error correction for the selected calibration class cannot be computed until you have measured all the necessary standards. |
31 |
ADDRESSED TO TALK WITH NOTHING TO SAY |
|
You have sent a read command to the analyzer (such as ENTER 716) without first requesting data with an appropriate output command (such as OUTPDATA). The analyzer has no data in the output queue to satisfy the request. |
212 |
ANALOG BUS DISABLED IN 6KHZ IF BW |
|
When you press AVG IF BW[6000] , the analog bus is disabled and not available for use in troubleshooting. |
60 |
ANALOG INPUT OVERLOAD |
|
The power level of the analog input is too high. Reduce the power level of the analog input source. |
183 |
BATTERY FAILED. STATE MEMORY CLEARED |
|
The battery protection of the non-volatile CMOS memory has failed. The CMOS memory has been cleared. |
184 |
BATTERY LOW! STORE SAVE REGS TO DISK |
|
The battery protection of the non-volatile CMOS memory is in danger of failing. If this occurs, all of the instrument state registers stored in CMOS memory will be lost. |
34 |
BLOCK INPUT ERROR |
|
The analyzer did not receive a complete data transmission. This is usually caused by an interruption of the bus transaction. Clear by pressing the “Local” key or aborting the I/O process at the controller. |
35 |
BLOCK INPUT LENGTH ERROR |
|
The length of the header received by the analyzer did not agree with the size of the internal array block. Refer to the analyzer’s programmers guide for instructions on using analyzer input commands. |
74 |
CALIBRATION ABORTED |
|
You have changed the active channel during a calibration so the calibration in progress was terminated. Make sure the appropriate channel is active and restart the calibration. |
63 |
CALIBRATION REQUIRED |
|
A calibration set could not be found that matched the current stimulus state or measurement parameter. You will have to perform a new calibration. |
203 |
CANNOT READ/WRITE 1 FILE SYSTEM |
|
The disk is being accessed by the analyzer and is found to contain an 1 (hierarchical file system) or files nested within subdirectories. The analyzer does not support 1. Replace the disk medium with a LIF or DOS formatted disk that does not contain files nested within subdirectories. |
179 |
CAUTION: POWER OUT MAY BE UNLEVELED |
|
There is either a hardware failure in the source or you have attempted to set the power level too high. The analyzer allows the output power to be set higher or lower than the specified available power range. However, these output powers may be un-leveled or unavailable. |
3 |
CORRECTION CONSTANTS NOT STORED |
|
A store operation to the EEPROM was not successful. You must change the position of the jumper on the A7 CPU assembly. |
66 |
CORRECTION TURNED OFF |
|
Critical parameters in your current instrument state do not match the parameters for the calibration set, therefore correction has been turned off. The critical instrument state parameters are sweep type, start frequency, frequency span, and number of points. |
64 |
CURRENT PARAMETER NOT IN CAL SET |
|
Correction is not valid for your selected measurement parameter. Either change the measurement parameters or perform a new calibration. |
11 |
DEADLOCK |
|
A fatal firmware error occurred before instrument preset completed. Call for service. |
119 |
DEVICE: not on, not connect, wrong addrs |
|
The device at the selected address cannot be accessed by the analyzer. Verify that the device is switched on, and check the GPIB connection between the analyzer and the device. Ensure that the device address recognized by the analyzer matches the GPIB address set on the device itself. |
39 |
DISK HARDWARE PROBLEM |
|
The disk drive is not responding correctly. If using an external disk drive, refer to the disk drive operating manual. |
190 |
DISK MESSAGE LENGTH ERROR |
|
The analyzer and the external disk drive aren’t communicating properly. Check the GPIB connection and then try substituting another disk drive to isolate the problem instrument. |
38 |
DISK: not on, not connected, wrong addrs |
|
The disk cannot be accessed by the analyzer. Verify power to the disk drive, and check the GPIB connection between the analyzer and the disk drive. Ensure that the disk drive address recognized by the analyzer matches the GPIB address set on the disk drive itself. |
189 |
DISK READ/WRITE ERROR |
|
There may be a problem with your disk. Try a new floppy disk. If a new floppy disk does not eliminate the error, suspect hardware problems. |
|
FILE NOT COMPATIBLE WITH INSTRUMENT |
|
You cannot recall user graphics that had been saved on an earlier model of analyzer with a monochrome display. |
46 |
ILLEGAL UNIT OR VOLUME NUMBER |
|
The disk unit or volume number set in the analyzer is not valid. Refer to the disk drive operating manual. |
47 |
INITIALIZATION FAILED |
|
The disk initialization failed, probably because the disk is damaged. |
154 |
INSUFFICIENT MEMORY, PWR MTR CAL OFF |
|
There is not enough memory space for the power meter calibration array. Increase the available memory by clearing one or more save/recall registers, or by reducing the number of points. |
69 |
NO CALIBRATION CURRENTLY IN PROGRESS |
|
The softkey “RESUME CAL SEQUENCE” is not valid unless a calibration is already in progress. Start a new calibration. |
114 |
NO FAIL FOUND |
|
Service Error, the self-diagnose function of the instrument operates on an internal test failure. At this time, no failure has been detected. |
45 |
NO FILE(S) FOUND ON DISK |
|
No files of the type created by an analyzer store operation were found on the disk. If you requested a specific file title, that file was not found on the disk. |
5 |
NO IF FOUND: CHECK R INPUT LEVEL |
|
The first IF signal was not detected during pretune. Check the front panel R channel jumper. |
70 |
NO SPACE FOR NEW CAL.CLEAR REGISTERS |
|
You cannot store a calibration set due to insufficient memory. You can free more memory by clearing a saved instrument state from an internal register (which may also delete an associated calibration set, if all the instrument states using the calibration kit have been deleted). You can store the saved instrument state and calibration set to a disk before clearing them. After deleting the instrument states, press “Preset” to run the memory packer. |
198 |
NOT ALLOWED DURING POWER METER CAL |
|
When the analyzer is performing a power meter calibration, the GPIB bus is unavailable for other functions such as printing or plotting. |
44 |
NOT ENOUGH SPACE ON DISK FOR STORE |
|
The store operation will overflow the available disk space. Insert a new disk or purge files to create free disk space. |
58 |
OVERLOAD ON INPUT A, POWER REDUCED |
|
You have exceeded approximately +14 dBm at one of the test ports. The RF output power is automatically reduced to -85 dBm. The annotation Pß appears in the left margin of the display to indicate that the power trip function has been activated. When this occurs, reset the power to a lower level, then toggle the “SOURCE PWR on OFF” softkey to switch on the power again. |
59 |
OVERLOAD ON INPUT B, POWER REDUCED |
|
You have exceeded approximately +14 dBm at one of the test ports. The RF output power is automatically reduced to -85 dBm. The annotation Pß appears in the left margin of the display to indicate that the power trip function has been activated. When this occurs, reset the power to a lower level, then toggle the “SOURCE PWR on OFF” softkey to switch on the power again. |
167 |
PARALLEL PORT NOT AVAILABLE FOR COPY |
|
You have defined the parallel port as general purpose I/O (GPIO) for sequencing. The definition was made under the “Local” key menus. To access the parallel port for copy, set the selection to “PARALLEL [COPY]”. |
165 |
PARALLEL PORT NOT AVAILABLE FOR GPIO |
|
You have defined the parallel port as COPY for sequencing in the GPIB menu. To access the parallel port for general purpose I/O (GPIO), set the selection to “PARALLEL [GPIO]”. |
4 |
PHASE LOCK CAL FAILED |
|
An internal phase lock calibration routine is automatically executed at power-on, preset, and any time a loss of phase lock is detected. This message indicates that phase lock calibration was initiated and the first IF detected, but a problem prevented the calibration from completing successfully. This message may appear if you connect a mixer between the RF output and R input before turning on frequency offset mode. Ignore it: it will go away when you turn on frequency offset. This message may also appear if you turn on frequency offset mode before you define the offset. |
7 |
PHASE LOCK FAILURE |
|
The first IF signal was detected at pretune, but phase lock could not be acquired. |
8 |
PHASE LOCK LOST |
|
Phase lock was acquired but then lost. |
6 |
POSSIBLE FALSE LOCK |
|
Phase lock has been achieved, but the source may be phase locked to the wrong harmonic of the synthesizer. |
116 |
POWER METER INVALID |
|
The power meter indicates an out-of-range condition. Check the test setup. |
118 |
POWER METER NOT SETTLED |
|
Sequential power meter readings are not consistent. Verify that the equipment is set up correctly. If so, preset the instrument and restart the operation. |
21 |
POWER SUPPLY HOT! |
|
The temperature sensors on the A8 post-regulator assembly have detected an over-temperature condition. The power supplies regulated on the post-regulator have been shut down. |
22 |
POWER SUPPLY SHUT DOWN! |
|
One or more supplies on the A8 post-regulator assembly have been shut down due to an over-current, over-voltage, or under-voltage condition. |
175 |
PRINTER: error |
|
The parallel port printer is malfunctioning. The analyzer cannot complete the copy function. |
177 |
PRINTER: not handshaking |
|
The printer at the parallel port is not responding. |
24 |
PRINTER: not on, not connected, wrong addrs |
|
The printer does not respond to control. Verify power to the printer, and check the GPIB connection between the analyzer and the printer. Ensure that the printer address recognized by the analyzer matches the GPIB address set on the printer itself. |
117 |
PWR MTR: NOT ON/CONNECTED OR WRONG ADDR |
|
The power meter cannot be accessed by the analyzer. Verify that the power meter address and model number set in the analyzer match the address and model number of the actual power meter. |
151 |
SAVE FAILED.INSUFFICIENT MEMORY |
|
You cannot store an instrument state in an internal register due to insufficient memory. Increase the available memory by clearing one or more save/recall registers and pressing “Preset”, or by storing files to a disk. |
112 |
SELF TEST #n FAILED |
|
Internal test #n has failed. Several internal test routines are executed at instrument preset. The analyzer reports the first failure detected. Refer to the internal tests and the self-diagnose feature descriptions earlier in this chapter. |
|
SOURCE POWER TURNED OFF, RESET UNDER POWER MENU |
|
You have exceeded the maximum power level at one of the inputs and power has been automatically reduced. The annotation Pß indicates that power trip has been activated. When this occurs, reset the power and then press “Power” “SOURCE PWR on OFF” to switch on the power. This message follows error numbers 57, 58, and 59. |
187 |
SWEEP MODE CHANGED TO CW TIME SWEEP |
|
If you select external source auto or manual instrument mode and you do not also select CW mode, the analyzer is automatically switched to CW. |
12 |
SWEEP TIME TOO FAST |
|
The fractional-N and digital IF circuits have lost synchronization. |
113 |
TEST ABORTED |
|
You have prematurely stopped a service test. |
57 |
TEST PORT OVERLOAD, REDUCE POWER |
|
You have exceeded approximately +14 dBm at one of the test ports (or 0 dBm at the A or B sampler, Option 012 only). When this occurs, reduce the power to a lower level. |
115 |
TROUBLE!CHECK SET-UP AND START OVER |
|
Service Error, your equipment setup for the adjustment procedure in progress is not correct. |
77 |
WRONG DISK FORMAT, INITIALIZE DISK |
|
You have attempted to store, load, or read file titles, but your disk format does not conform to the Logical Interchange Format (LIF) or DOS format. You must initialize the disk before reading or writing to it. |