The Agilent 83496B Clock Recovery Module with Phase Noise Analysis provides increased eye-mask and jitter measurement accuracy with breakthrough performance in clock recovery circuitry. It provides ideal performance for waveform analysis with the 86100 Digital Communications Analyzer and can derive a clock from signals with rates as low as 50 Mb/s-14.2 GB/s, and any rate in between. At under 300 femtoseconds rms, the residual jitter of the output clock is virtually negligible, allowing accurate measurements of very low levels of signal jitter. Specifications. Operating range: 50 Mb/s to 7.1 Gb/s(with Option 200: 50 Mb/s to 14.2 Gb/s). Residual jitter (jitter free input): < 300 fs rms. Loop BW: 270 kHz or 1.5 MHz (with Option 300: 15 kHz to 10 MHz or fixed ratio of data rate/N). Wavelength range: 780 to 1330 nm MMF (62.5 um), Option 101 1250 to 1620 nm SMF (9 um). Options. Option 100: Electrical differential or single-ended clock recovery 50 Mb/s to 7.1 Gb/s. Option 101: Optical single mode 1250 to 1620 nm and multimode 780 to 1330 nm clock recovery 50 Mb/s to 7.1 Gb/s. Option 200: Increase operating range to 14.2 Gb/s. Option 201: Operating range of either Option 100 or Option 101 is set to be 7.1 to 14.2 Gb/s. Option 300: Add Golden PLL (tunable loop bandwidth capability).
|