The HP 54542A is a 500 MHz, 2 GSa/s sample rate, 4-channel digitizing oscilloscope. It has all of the features that you would expect in a scope plus more. Use sequential single-shot when you need to capture successive single-shot events without capturing the dead time in between. It is a great tool for applications such as pulsed laser research, high-energy physics, and pulse echo. Use glitch trigger to find the causes of anomalies in circuit operation. Trigger on hard-to-see narrow glitches down to 1 ns wide. Waveforms can be saved and time-stamped upon failure of either wave form compare or limit test. Up to 665 failures of 500 points each can be stored in multiple memory and sent to a printer, plotter, or a computer via HP-IB. Specifications. Repetitive Bandwidth: 500 MHz (equivalent time). Real-time bandwidth: 500 MHz (all channels). Number of channels (all are simultaneous acquisition): 4. Vertical sensitivity range: 1 mV/div to 5 V/div. Vertical dc gain Accuracy: ± 1.25% of full scale. Vertical resolution: 8 bits (10 bits via HP-IB with averaging). Maximum sample rate (Real time mode): 2 GSa/s (all channels). Maximum sample rate (Repetitive mode): 1 GSa/s. Waveform record length: 32,768 points (real-time), 501 points (repetitive). Maximum input: 1 M-ohm: ± 250 V (dc + ac) [ac<10 kHz], 50 ohm: 5 V rms. Time base range: 500 ps/div to 5 s/div. Resolution: 10 ps. Trigger Pulse width (minimum) 1 ns. Monochrome Display. Opt 001 Telecommunications Mask Test Software.
|