The HP 70950A is a 1700 nm Optical Spectrum Analyzer Module for 70000 Modular Measurement System (MMS). It is a 1/2 width module with FC/PC connectors standard. Sensitivity is -80 dBm and dynamic range is 60 dB 1 nm away from the signal. The resolution is 0.1 to 10 nm in a 1,2,5 sequence. The input connector system has a wide choice of adapter options. It offers three advanced functions to measure and characterize LED, DFB, or Fabry-Perot lasers. Specifications. Wavelength Range: 600 to 1700 nm. Wavelength Tuning repeatability: ±0.005 nm. Wavelength Span Range: 0.2 nm to full range. Wavelength Absolute Accuracy: ±1 nm. Maximum Sensitivity (1100-1600 nm): –90 dBm.Sweep Time sensitivity in auto, full span: < 0.5 seconds. Input Power, 1 dB compression level within resolution: =10 dBm. Dynamic Range: -50 dB at = ± 1 nm(600 to 1700 nm), -55 dB at = ± 0.5 nm(1250 to 1600 nm). Resolution: 0.08 nm, 0.1 nm, 10 nm (in a 1,2,5 sequence). Options. 001, Programmable Current Source. 002, Built-in White Light Source. 010, Delete FC/PC Interface. 051, EDFA Test Personality. 101, Enhanced Polarization Sensitivity.
|