600 nm to 1700 nm, Optical Spectrum Analyzer Module for 70000 Modular Measurement System (MMS). 1/2 width module with FC/PC connectors standard. Wavelength Tuning repeatability: ±0.005 nm. Wavelength Span Range: 0.2 nm to full range and zero span. Wavelength Absolute Accuracy: ±1 nm. Flatness: 1530 nm - 1570 nm, ±0.25 dB; 1250 nm - 1600 nm, ±1 dB. Maximum Sensitivity (1100-1600 nm): –90 dBm. Maximum Sweep Rate: 40 nm/50 ms. Max Measurement Power: + 20 dBm per 5 nm, +30 dBm total. Dynamic Range in 0.1 nm resolution: -50 dB at = ± 1 nm. Resolution Bandwidth: 0.08 nm, 0.1 nm, 0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm, 10 nm.
|