The Agilent 8481B high power sensor measures average power over the frequency range 10 MHz to 18 GHz and power range -0 to +44 dBm. It is designed for use with an Agilent / HP 435B, 436A, 437B, 438A, 70100A power meters. It is also compatible with EPM, EPM-P and P-series power meters, plus the E1416A VXI. Specifications. Frequency range: 10 MHz to 18 GHz. Maximum SWR: 10 MHz to 2 GHz, 1.10; 2 to 12.4 GHz, 1.18; 12.4 to 18 GHz, 1.28. Power Linearity: +35 to +44 dBm; ±4%. Maximum power: 0° to 35° C: 30 W avg, 35° to 55° C: 25 W avg, 0.01 to 5.8 GHz: 500 W pk, 5.8 to 18 GHz: 125 W pk, 500 W-µs per pulse. Connector type: Type-N(m) 50 ohm.
|