The 86100B Infiniium DCA can be viewed as three high-performance instruments in one. It’s a general-purpose wide-bandwidth sampling oscilloscope, a digital communications analyzer and a time domain reflectometer. Standard Measurements; Oscilloscope Mode (Time): Rise Time, Fall Time, Jitter RMS, Jitter p-p, Period, Frequency, + Pulse Width, - Pulse Width, Duty Cycle, Delta Time, [Tmax, Tmin, Tedge—remote commands only]. Oscilloscope Mode (Amplitude): Overshoot, Average Power, V amptd, V p-p, V rms, V top, V base, V max, V min, V avg. Eye/Mask Mode (NRZ Eye Measurements): Extinction Ratio, Jitter RMS, Jitter p-p, Average Power, Crossing Percentage, Rise Time, Fall Time, One Level, Zero Level, Eye Height, Eye Width, Signal to Noise (Q-Factor), Duty Cycle Distortion, Bit Rate, Eye Amplitude. Eye/Mask Mode (RZ Eye Measurements): Extinction Ratio, Jitter RMS, Jitter p-p, Average Power, Rise Time, Fall Time, One Level, Zero Level, Eye Height, Eye Amplitude, Opening Factor, Eye Width, Pulse Width, Signal to Noise (Q-Factor), Duty Cycle, Bit Rate, Contrast Ratio. Mask Test: Open Mask, Start Mask Test, Exit Mask Test, Filter, Mask Test Margins, Mask Test Scaling, Create NRZ Mask. TDR/TDT Mode (requires TDR module): Quick TDR, TDR/TDT Setup, Normalize, Response, Rise Time, Fall Time, delta Time. Time Base Scale Factor: Min 2 ps/div (with 86107A: 500 fs/div), Max 1 s/div. Time Base Delay: Min >=24 ns (Standard), 28 ns (Option 001). Number of Channels: 4 (simultaneous acquisition). Vertical Resolution: 12 bit A/D converter (up to 15 bits with averaging). Trigger Sensitivity: 200 m Vpp (sinusoidal input or 200 ps minimum pulse width). Trigger Bandwidth (Module dependent): 2.4 to 43 GHz.
|