Multi-Wavelength Meter. Wavelength : 700 nm to 1650 nm. Absolute accuracy (lines separated by >=30 GHz): ±3 ppm (±0.005 nm at 1550 nm and ±0.004 nm at 1310 nm). Differential accuracy (characteristic): ±2 ppm. Minimum resolvable separation (characteristic): 20 GHz (0.16 nm at 1550 nm, 0.11 nm at 1300 nm). Display resolution: 0.001 nm. Amplitude Linearity, 1200 nm to 1600 nm, lines above -20 dBm: ±0.3 dB. Flatness, ±30 nm from any wavelength: 1200-1600 nm (characteristic): ±0.2 dB. Display resolution: 0.01 dB. Maximum safe input level (sum of all lines): +18 dBm. Measure up to 100 laser lines simultaneously.
Repair Services / Additional Information.
Common 86120 Repairable Errors We have repaired over 1000 86120's with a success rate of over 99% and with most replacement parts in stock. Our typical repair time is less than 5 business days. The following is a list of common repairable errors (In order of prevalence). E14 (ERROR 14) "DATA ACQUISITION PROBLEM" or "DATA ACQUISITION ERROR" We repair the E14 error using NEW, original manufacturer parts, which comes with a one year warranty. We don't use refurbished or cheap generic parts. E4 (ERROR 4) "MOTOR INDEX PULSE NOT FOUND" E7 (ERROR 7) "MOTOR NOT SETTLED" E34 (ERROR 34) "BATTERY FAILED"
|