The Agilent E4416A single-channel EPM-P power meter is used with E9320 power sensors to make accurate power measurements. You can perform peak, peak-to-average ratio, and average power measurements on all the common modulation formats used in wireless communications. Time-gated power measurements can also be made using the meters’ comprehensive triggering features. It can perform power measurements at rates up to 1000 readings per second over the GPIB. The E9320 power sensor family has two frequency ranges are available: 50 MHz to 6 GHz, and 50 MHz to 18 GHz. There are also three modulation bandwidths: 300 kHz for TDMA signals, 1.5 MHz for CDMA (IS-95A) signals and 5 MHz for wideband-CDMA signals. Specifications. Frequency Range: 9 kHz to 110 GHz, sensor dependent. Power Range: –70 to +44 dBm, sensor dependent. Display Units Absolute: Watts or dBm; Relative: Percent or dB. Display Resolution: Selectable resolution of 1.0, 0.1, 0.01, 0.001 dB in logarithmic mode, or 1 to 4 significant digits in linear mode. Modulation Bandwidth: 5MHz maximum (set by meter and is sensor dependent). Time Base Accuracy: 0.1%. Trigger Sources: Internal, External TTL, GPIB, RS232/422. Options. Opt 002 Supplies rear-panel sensor input (power reference calibrator is on the front panel). Opt 003 Supplies rear-panel sensor input (power reference calibrator is on the rear panel).
|